Main unit 

 
 
 
 
 
 
 
 
 
 
 
 

  Display 

:  Displays measurement readings or status of the unit. 

  PWR key 

:  Press this key to turn on the unit. To turn off the unit, press the key again. 

  CAL key 

:  Enters the calibration mode. 

  SKIP key 

:  Moves the pH set values while in the calibration mode. 

Enters the manual temperature setting mode when pressed two seconds or 
more after the connecting plug for the temperature sensor is disconnected, 
while in the measurement mode, calibration mode or MAX/MIN logging mode. 

  REC key 

: The calibration value for the pH 4, pH 7 or pH 10 is stored in the unit memory, 

if pressed while in the calibration mode. 

Pressing this key while in the measurement mode will move to the MAX/MIN 
logging mode and start logging the MAX value (the highest pH value) and the 
MIN value (lowest pH value). Pressing the key again will alternately display the 
MAX value and MIN value. The maximum and minimum voltage (mV) can also be 
logged. To return to the measurement mode, press the REC key for three 
seconds or more while in the MAX/MIN logging mode. Pressing this key while in 
the manual temperature setting mode will enable the liquid temperature. 

  pH/mV key 

: Press this key to switch the pH and mV in the upper LCD in measurement 

mode. 

Increases (↑) the pH values to be calibrated in the calibration mode. 

Increases (↑) the set temperature value in the manual temperature 
measurement mode. 

             key 

(backlight key) 

: Press this key to turn on the backlight on the LCD in the measurement mode. 

Decreases (↓) the pH values to be calibrated in the calibration mode. 
Decreases (↓) the set temperature value in the manual temperature 
measurement mode. 

  Battery cover 

 

  Screw for battery cover                                     

 

 

Top face of the case 

 

   

  pH sensor connecting jack: Used to connect the sensor.     

            This product uses the BNC connector. 

  Temperature sensor connecting jack: Used to connect the     

temperature sensor. This product uses the 2.5-mm diode 
plug. 

 

10 

11 

Содержание SK-640PH

Страница 1: ...Instruction manual for No 6422 00 Handy type pH meter Model SK 640PH No 3 Nishi Fukudacho Kanda Chiyodaku Tokyo 101 0037 Japan URL http www sksato co jp english N 11...

Страница 2: ...e If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may increase Never disassemble or modify the unit Doing so may cau...

Страница 3: ...mperature compensation Features Large LCD for easy reading The pH value and the temperature are indicated on a large LCD Auto Power Off The auto power off function turns off the power automatically if...

Страница 4: ...e and MIN value The maximum and minimum voltage mV can also be logged To return to the measurement mode press the REC key for three seconds or more while in the MAX MIN logging mode Pressing this key...

Страница 5: ...minimum pH or voltage value is displayed BAT Lights up when battery power is insufficient ATC Lights up when the temperature sensor is connected Display of time voltage Indicates the time elapsed sin...

Страница 6: ...g the vinyl cover disconnect the sensor probe first and then remove in the reverse order of installation Note The vinyl cover is not waterproof splash proof according to the Japanese Industrial Standa...

Страница 7: ...on the right 2 Installing the glass electrode protection bottle 2 1 Install the protection bottle to the sensor probe in the reverse order of removal 2 2 Hold the sensor probe upright so as not to spi...

Страница 8: ...fter replacing the battery c When very accurate measurement is necessary For more accurate measurement calibration of about once a day is recommended For calibrating the SK 620PH prepare the following...

Страница 9: ...alibration value is 4 04 7 Press the REC key to store the calibration value in the unit memory SA is displayed on the LCD to indicate that the calibration data has been stored in the unit memory Now t...

Страница 10: ...ssible to perform calibration using the oxalate standard solution pH 1 68 at 25 C or borate standard solution pH 9 18 at 25 C CAL key REC key REC Key REC Key Measurement mode Calibration pH4 Calibrati...

Страница 11: ...hich is capable of measuring both pH and temperature is used connect both the BNC plug pH and the diode plug temperature 2 Press the PWR key to turn on the unit The whole LCD lights up for approximate...

Страница 12: ...21...

Страница 13: ...eft unwashed the liquid measured tends to form a film on the surface of the electrode causing measurement errors Always wash the pH electrode every time after use Washing can be done with water distil...

Страница 14: ...bout the RS 232C output RS 232C output This instrument outputs data in real time on the pH voltage and temperature while in the measurement mode or the MAX MIN logging mode When the unit is connected...

Страница 15: ...d replacing battery The power to the unit is turned off The auto power off is turned on Cancel the auto power off on the main unit Before measuring Error code pH display section The sensor probe is no...

Страница 16: ...an electric conductivity of approx 100 S cm or more The temperature of the sample largely fluctuated Make the sample temperature stable The pH values change with temperature Readings are abnormal The...

Страница 17: ...unication conditions or protocols are not correct Check the communication conditions or protocols About the RS 232C output The serial port on the computer is not open Check that the serial port is ope...

Страница 18: ...ent Temperature 0 to 50 Humidity lower than 80 rh no condensing Battery Life About 20 hours continuous Materials Case ABS resin Electrode Glass Probe Epoxy resin Dimensions Main unit W 72 H 182 D 31 5...

Страница 19: ...28 6400 40 PH40 Carrying case 6400 50 PH50 Electrode protection bottle...

Отзывы: