SATO SK-640PH Скачать руководство пользователя страница 13

22 

 

About the glass electrode 

 

(1)

 Storage

 

 

If left unused for a long time, the glass electrode becomes too dry to measure the pH accurately. 

 

The pH electrode works best and shows accurate readings when used with moderate moisture. 
When using the instrument for the first time after purchasing or a long storage period, we 
recommend placing the glass electrode in water (distilled, pure or tap water) or the pH standard 
solution for at least 12 hours and performing calibration before measuring. For everyday storage, 
fit the glass electrode protection bottle to the sensor probe. The glass electrode protection bottle is 
filled with the phthalate standard solution (pH 4) at the factory. 

 
(2) 

Inspection and calibration

 

 

For an accurate pH measurement, perform calibration using the pH standard solution. 

 

If there is a deviation from the expected value, adjust it before performing a pH measurement. 

 
 

Note that there are cases where the pH value of the standard solution has changed from its 
original value. This can happen if the pH standard solution is dirty or if storing the solution for a 
long time. It is advisable to use a new pH standard solution, where applicable. 

 
(3) 

Washing 

 

Be sure to wash the pH electrode after each pH calibration or measurement with water (distilled, 
pure or tap water). If left unwashed, the liquid measured tends to form a film on the surface of the 
electrode, causing measurement errors. Always wash the pH electrode every time after use. 

 

Washing can be done with water (distilled, pure or tap water) in most cases, but if there is a 
stubborn stain, use neutral detergent diluted in water to wash only the glass electrode section 
(spherical part). After using neutral detergent, thoroughly rinse the electrode with water (distilled, 
pure or tap water) to remove the neutral detergent, and then immerse the electrode in the pH 
standard solution (pH 4) before measurement. 

 
 

Note that neutral detergent cannot be used on the liquid junction of the sensor probe. Otherwise, 
the performance of the inner solution could be damaged if the neutral detergent penetrates inside 
the sensor. 

 
(4) 

Useful life

 

 

It is not possible to replenish or repair the inner solution of the sensor probe. 

 

The pH electrode is a consumable. Its useful life varies depending on the liquid to be measured, 
pH value or temperature. 

 
(5) 

Deposit of white potassium chloride crystals 

 

A deposit of white crystals is sometimes found on the tip of the electrode or protection bottle; this 
does not pose a problem when using the instrument. The deposit is easily washed off with tap 
water. 

 

If it does not come off with tap water, soak the electrode in warm water to remove it. 

 

Do not lick or swallow the potassium chloride crystals. If the eyes or skin becomes contaminated 
with the crystals, immediately consult a doctor. 

Содержание SK-640PH

Страница 1: ...Instruction manual for No 6422 00 Handy type pH meter Model SK 640PH No 3 Nishi Fukudacho Kanda Chiyodaku Tokyo 101 0037 Japan URL http www sksato co jp english N 11...

Страница 2: ...e If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may increase Never disassemble or modify the unit Doing so may cau...

Страница 3: ...mperature compensation Features Large LCD for easy reading The pH value and the temperature are indicated on a large LCD Auto Power Off The auto power off function turns off the power automatically if...

Страница 4: ...e and MIN value The maximum and minimum voltage mV can also be logged To return to the measurement mode press the REC key for three seconds or more while in the MAX MIN logging mode Pressing this key...

Страница 5: ...minimum pH or voltage value is displayed BAT Lights up when battery power is insufficient ATC Lights up when the temperature sensor is connected Display of time voltage Indicates the time elapsed sin...

Страница 6: ...g the vinyl cover disconnect the sensor probe first and then remove in the reverse order of installation Note The vinyl cover is not waterproof splash proof according to the Japanese Industrial Standa...

Страница 7: ...on the right 2 Installing the glass electrode protection bottle 2 1 Install the protection bottle to the sensor probe in the reverse order of removal 2 2 Hold the sensor probe upright so as not to spi...

Страница 8: ...fter replacing the battery c When very accurate measurement is necessary For more accurate measurement calibration of about once a day is recommended For calibrating the SK 620PH prepare the following...

Страница 9: ...alibration value is 4 04 7 Press the REC key to store the calibration value in the unit memory SA is displayed on the LCD to indicate that the calibration data has been stored in the unit memory Now t...

Страница 10: ...ssible to perform calibration using the oxalate standard solution pH 1 68 at 25 C or borate standard solution pH 9 18 at 25 C CAL key REC key REC Key REC Key Measurement mode Calibration pH4 Calibrati...

Страница 11: ...hich is capable of measuring both pH and temperature is used connect both the BNC plug pH and the diode plug temperature 2 Press the PWR key to turn on the unit The whole LCD lights up for approximate...

Страница 12: ...21...

Страница 13: ...eft unwashed the liquid measured tends to form a film on the surface of the electrode causing measurement errors Always wash the pH electrode every time after use Washing can be done with water distil...

Страница 14: ...bout the RS 232C output RS 232C output This instrument outputs data in real time on the pH voltage and temperature while in the measurement mode or the MAX MIN logging mode When the unit is connected...

Страница 15: ...d replacing battery The power to the unit is turned off The auto power off is turned on Cancel the auto power off on the main unit Before measuring Error code pH display section The sensor probe is no...

Страница 16: ...an electric conductivity of approx 100 S cm or more The temperature of the sample largely fluctuated Make the sample temperature stable The pH values change with temperature Readings are abnormal The...

Страница 17: ...unication conditions or protocols are not correct Check the communication conditions or protocols About the RS 232C output The serial port on the computer is not open Check that the serial port is ope...

Страница 18: ...ent Temperature 0 to 50 Humidity lower than 80 rh no condensing Battery Life About 20 hours continuous Materials Case ABS resin Electrode Glass Probe Epoxy resin Dimensions Main unit W 72 H 182 D 31 5...

Страница 19: ...28 6400 40 PH40 Carrying case 6400 50 PH50 Electrode protection bottle...

Отзывы: