4.
Narrative
Description
‐
Connect
ZCM17
with
hose
between
the
metering
system
to
be
tested
and
the
receiving
tank
‐
Start
a
delivery
to
fill
the
ZCM17
up
to
the
transfer
point
‐
Reset
the
mechanical
head
/
Put
the
Equalis
S
in
delivery
mode
‐
Start
delivery
with
a
preset
volume
‐
Compare
the
volume
displayed
by
the
master
meter
with
the
one
displayed
by
the
meter
to
be
tested
‐
Adjust
the
meter
correction
coefficient
‐
Start
a
new
delivery
to
confirm
the
correction
‐
Use
the
drainage
hand
pump
to
empty
the
ZCM17
to
limit
pollution
when
disconnecting
or
testing
with
a
new
fluid
‐
Disconnect
the
ZCM17
hose
5.
Periodic
Servicing
As
a
general
rule,
a
periodic
maintenance
of
at
least
once
a
year
is
recommended.
Service
operations
must
be
carried
out
by
a
Weights
&
Measures
approved
Company.
6.
Cleaning
We
strongly
advise
against
the
use
of
a
high
pressure
water
jet
to
clean
the
measuring
unit,
as
this
could
seriously
damage
the
metering
unit.
Содержание ZCM17
Страница 1: ...02 2018 Edition 1 19 U518156 en REV0 MASTER METER Description Operation Maintenance ...
Страница 12: ...02 2018 Edition 12 19 U518156 en REV0 B Adjustment Procedure with EQUALIS S ...
Страница 13: ...02 2018 Edition 13 19 U518156 en REV0 ...
Страница 14: ...02 2018 Edition 14 19 U518156 en REV0 ...
Страница 15: ...02 2018 Edition 15 19 U518156 en REV0 ...