![Sartorius VIS1X Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/sartorius/vis1x/vis1x_operating-instruction_1199693054.webp)
54
VIS1X
Operating Instructions
Technical Data
15.2 Model-specific Data
Specification
Unit
Value
Weighing capacity
g
7500 | 999.95
Readability
g
0.1 | 0.05
Tare range (subtractive)
g
-7500
External adjustment weight/
accuracy class
kg
1, 2, 5 /
F2 or better
Diameter of weighing pan
mm
233
Net weight
kg
2.4
15.3 Verified Models with EC Type Examination Certificate: Model-specific Technical Data
Specification
Unit
Value
Accuracy class
II
Type
PMA-EV
Weighing capacity max.
g
7500
Weighing capacity min.
g
5
Scale interval d
g
0.1
Verification scale interval e
g
1
Number of scale Intervals n
7500
Temperature range
°C
+10 to +30
Tare equalization range (subtractive)
<
100% from max. weighing capacity
Nominal load of the load receptor *
g
8000
Switch-on zero setting
g
±
375
Diameter of weighing pan
mm
233
* The sum of the max. switch-on zero setting and dead load may not exceeded the nominal load of the load receptor.
Содержание VIS1X
Страница 2: ......
Страница 58: ...58 VIS1X Operating Instructions Conformity Approvals...
Страница 72: ...72 VIS1X Operating Instructions Conformity Approvals...
Страница 73: ...VIS1X Operating Instructions 73 Conformity Approvals...
Страница 74: ...74 VIS1X Operating Instructions Conformity Approvals...
Страница 75: ...VIS1X Operating Instructions 75 Conformity Approvals 18 3 EU Type Examination Certificate Verified Models...