Place empty container on
the platform
)
1. Tare the scale.
Note: If the automatic tare
function is enabled, you do
not need to press the
)
key
to tare the scale; the tare
weight is saved automatically
when you place the container
on the platform.
2. Place 1
st
animal in container
20
3. Enter the number of
subweighing operations using
the keypad (in this example,
20 measurements)
r
4. Save the value entered and
begin averaging
The averaging routine does
not begin until the fluctuation
in weight value remains below
a defined threshold over three
consecutive measurements.
The number of subweighing
operations remaining is shown
in the numeric display.
Read the result of averaging
p
5. Print the results.
Note: If automatic printout
of results is enabled, you do
not need to press the
p
key;
the results are printed auto-
matically.
mDef
+
20
Configured printout:
see page 82
T
+
0.292 kg
x-Net +
0.183 kg
--------------------
When you unload the
weighing platform, the display
switches to the weight readout
automatically, unless con-
figured otherwise in the oper-
ating menu. The weighing
instrument is ready for the
next measurement.
66
Example:
Measuring the weight of one mouse.
Settings (changes in the factory settings required for this example):
Setup: Application: Animal weighing
Setup: PRTPROT (printout): 7.7.x (COM1) (see “Configuration” for options)
Содержание Midrics 1
Страница 1: ...Operating Instructions Sartorius Midrics 1 Midrics 2 Models MIS1 MIS2 Indicators 98648 014 89 ...
Страница 92: ...92 Dimensions Scale Drawings All dimensions given in millimeters 92 ...
Страница 97: ...97 C ...
Страница 98: ...98 ...
Страница 99: ...99 ...
Страница 100: ...100 ...
Страница 101: ...101 ...
Страница 102: ...102 ...
Страница 104: ...104 ...