Adjusting the Analyzer
Heating Element Adjustment
The procedure for adjusting the heating
element is described in the instructions
supplied with the YTM15MA temperature
adjustment set .
Weighing System Adjustment
To adjust the weighing system, perform
calibration and adjustment as described
in the following .
Purpose
Calibration is the determination of the
difference between the weight readout
and the true weight (mass) of a sample .
Calibration does not entail making any
changes within the weighing system .
Adjustment is the correction of the
difference between the measured value
displayed and the true weight (mass)
of a sample, or the reduction of the
difference to a level within specified
permissible error limits .
Features
Calibration is performed externally with
the following weight value:
– MA35: 30 g; see “Accessories,” order no .
YSS43-02
You can have calibration and adjust-
ment results documented as a ISO/GLP-
compliant printout (see the page after
next for an example) .