Setting the Drying Parameters
Select PRG in the function line to adapt parameters for the drying program to the particular
requirements of the product sampled .
Drying Parameters
Temperature during heating
40 … 160°C
End of analysis
0 .0 min
Select 0 .0 minutes for fully automatic shutoff
0 .1 to 99 min
Select an interval from 0 .1 to 99 minutes to define a specific
analysis time
Display mode for result
%M
Moisture
%S
Dry weight
%MS
Ratio
g
Residual weight
Start of analysis
E
With stability, after the
key is pressed
A
Without stability, after the cover is closed
Print intermediate results
0 .0 min
Off
0 .1 to 10 .0 min