85
Overview
Verified Models with EC Type Approval
General Specifications
AC power source/power requirements
AC adapter, 230 or 115 V, + 15%... –20%
Frequency
48 – 60 Hz
Dust and water protection rating
according to EN 60529
1
)
IP54 (protected against harmful dust deposits and splashes of water)
Adaptation to ambient conditions
By selection of 1 of 4 optimized filter levels
Display update
0.1 – 0.4 (depends on the filter level selected)
Power consumption
16 VA: maximum; 9 VA: average
Hours of operation with fully charged
YRB 06 Z external battery pack, approx.
14 h
Selectable weight units
Grams, kilograms
Selectable application programs
Mass unit conversion, counting, weighing in percent, net-total formulation, animal weighing,
over/under checkweighing, recalculation
Built-in interface
RS-232 C
Format:
7-bit ASCII, 1 start bit, 1 or 2 stop bits
Parity:
odd, even, or space
Transmission rates:
150 to 19,200 baud
Handshake:
Software or hardware mode
1
) specially protected dust-tight and washdown resistant AC adapter; see the section on “Accessories”
Specifications of the Individual Models
Model
LP1200S-0CE
LP620S-0CE
LP220S-0CE
LP620P-0CE
Type
isoTEST in conjunction with BD BF
Accuracy class*
k
K
K
K
Scale interval, d*
g
0.001
0.001
0.001
0.001/0.002/
0.005
Maximum weighing capacity, Max.*
g
1,200
620
220
120/240/620
Verification scale interval, e*
g
0.01
0.01
0.01
0.01
Minimum capacity, Min.*
g
0.1
0.02
0.02
0.02
Tare range (subtractive)
≤
100% of the max. weighing capacity
Application range according to CD*
g
0.1 – 1,200
0.02 – 620
0.02 – 220
0.02 – 620
Response time (average)
s
1.5
Allowable operating temperature range
0 ... +40
°
C (273 ... 313 K, 32
°
F ... 104
°
F) with the isoCAL function
1
)
External calibration weight value
g
1,000 (E2)
–
–
–
(of at least accuracy class...)
Other permissible external calibration
–
–
–
–
weights (of at least accuracy class)
Selectable weight units
Grams and kilograms
Pan size
mm
Ø 130
Dimensions (W x D x H)
mm
240x360x147
Net weight, approx.
kg
8.3
6.9
6.9
6.9
1
) = With the isoCAL function deactivated, the verified balance can only be used within the limited temperature range (can only be modified
by the Sartorius Service Center):
For balances of accuracy class
k
: +15
°
C to +25
°
C (59
°
F to 77
°
F)
For balances of accuracy class
K
: +10
°
C to +30
°
C (50
°
F to 86
°
F)
*
CD = Council Directive 90/384/EEC for non-automatic weighing instruments used within the European Economic Area
Содержание LP Models
Страница 18: ...18 setting not applicable in balances verified for use in legal metrology...
Страница 20: ...20 not in balance models with a weighing capacity 16 kg...
Страница 81: ...81 Overview Menu Structure setting cannot be changed in balances used for legal metrology...
Страница 92: ...92...
Страница 93: ...93 Overview...
Страница 94: ...94...
Страница 95: ...95 Overview Z 14 755 320 34 10 93...
Страница 96: ...96...