82
Specifications
General Specifications
AC power source/power requirements
AC adapter, 230 or 115 V, +15%… – 20%
Frequency
48 – 60 Hz
Allowable ambient operating temperature
0…+40
°
C (273…313 K, 32
°
F…104
°
F)
Operating temperature range
+ 10…+ 30
°
C
Dust and water protection rating
according to EN 60529*
IP54 (protected against harmful dust deposits and splashes of water)
Adaptation to ambient conditions
By selection of 1 of 4 optimized filter levels
Display update
0.1 – 0.4
(depends on the filter level selected)
Power consumption
16 VA: maximum; 9 VA: average
Hours of operation with fully charged
YRB 06 Z external battery pack, approx.
14 h
Selectable weight units
Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels, Singapore taels,
Taiwanese taels, grains, pennyweights, milligrams, parts per pound, Chinese taels, mommes,
Austrian carats, tola, baht and mesghal
Selectable application programs
Mass unit conversion, counting, weighing in percent, net-total formulation,
animal weighing, over/under checkweighing, recalculation
Built-in interface
RS-232 C
Format:
7-bit ASCII, 1 start bit, 1 or 2 stop bits
Parity:
odd, even or space
Transmission rates:
150 to 19,200 baud
Handshake:
Software or hardware mode
* = specially protected dust-tight and washdown-resistant AC adapter; see the section on “Accessories”
Specifications of the Individual Models
Model
LP1200S
LP620S
LP220S
LP620P
LP3200D
Readability
g
0.001
0.001
0.001
0.001/0.002/
0.001/0.01
0.005
Weighing capacity
g
1,200
620
220
120/240/620
1,010/3,200
Tare range (subtractive)
g
– 1,200
– 620
– 220
– 620
– 3,200
Repeatability (standard deviation)
≤±
g
0.001
0.001
0.001
0.001/0.001/
0.001/0.01
0.003
Linearity
≤±
g
0.002
0.002
0.002
0.002/0.002/
0.002/0.01
0.005
Sensitivity drift 10…+30
°
C
≤±
/K
2 • 10
–6
Response time (average)
s
1.5
External calibration weight
g
1,000 (E2)
500 (E2)
200 (E2)
500 (F1)
1,000 (E2)
(of at least accuracy class...)
Other allowable external
g
–
300, 400,
100 (E2)
200, 300, 400,
2,000, 3,000 (E2)
calibration weights
600 (E2)
600 (F1)
(of at least accuracy class…)
Pan size
mm Ø 130
Dimensions (W x D x H)
mm
240x360x147
Net weight, approx.
kg
8.3
6.9
6.9
6.9
8.4
Содержание LP Models
Страница 18: ...18 setting not applicable in balances verified for use in legal metrology...
Страница 20: ...20 not in balance models with a weighing capacity 16 kg...
Страница 81: ...81 Overview Menu Structure setting cannot be changed in balances used for legal metrology...
Страница 92: ...92...
Страница 93: ...93 Overview...
Страница 94: ...94...
Страница 95: ...95 Overview Z 14 755 320 34 10 93...
Страница 96: ...96...