Time-Controlled Functions
K
Purpose
With this application program, you
can configure the scale to perform
certain functions (e.g., automatic print-
out of values, store value in totalization
memory) at a given time or after a set
interval.
You can use the “Time-Controlled
Functions“ application in combination
with a program chosen from Applica-
tion 1 (e.g., counting, weighing in
percent) and one from Application 3
(e.g., totalizing, formulation).
Available Features
– Time-controlled activation of scale
functions:
– one time only, at a given time
(
Setting=
is displayed in the
text line)
– repeatedly, at given intervals
(
Interval=
is displayed in
the text line before the function is
started, and
Repeat=
is displayed after the
function is started)
Functions that can be time-controlled
include:
– Acoustic signal
– Lock in readout
– Automatic printout of values
– Store values for totalizing,
formulation or statistics
– Print time in addition to weight value
– Store value independent of stability
– Tare the scale after printout of weight
values
– Press the corresponding soft key to
cancel time-controlled functions
Factory Settings
Function after time interval:
Automatic printout of
values
Automatic function restart:
On
Storage mode:
Without stability
Print then tare:
On
Soft Key Functions
Stop
Stop the application
Quit
Confirm performed function
(e.g., “Lock in readout” or
“Beep”)
Interv
Store input interval for time-
controlled functions
Set.
Store input time for one-time
performance of function
Printout for Time-Controlled Functions
If the “Automatic printout of values”
parameter is set, the time and weight
(or other value) are printed out.
Time:
10:15:00
N
+
150.00 g
Time:
Time that the values were
stored
N:
Net weight
77
Preparation
●
Turn on the scale: press
e
> The Sartorius logo is displayed
●
Select the “Time-controlled functions” application in the Setup menu: press
H
●
Select the
Application parameters
: press the
q
key 2
+
,
then the
O
soft key
●
Select
Application 2 (control functions)
: press the
q
soft key,
then the
O
soft key
●
Select
Time-controlled functions
: press the
Q
or
q
soft key
●
Confirm
Time-controlled functions
: press the
O
soft key
Time-controlled
Function after
Beep
functions
time interval
Lock in readout
o Automatic printout of values
Store value in applicat. 3 memory
Automatic function
o On
restart
Off
Storage mode
o Without stability
After stability
After higher stability
Print then tare
On
o Off
o = factory setting
see also the “Application Parameters (Overview)” in the chapter entitled “Configuration”
●
Save settings and exit the Setup menu: press the
oo
soft key
Print Net Values without Printout of Time
Select the Setup menu:
Setup: Printout: Application-defined output: Auto print upon initialization: Off