![Sartorius Factory Series FC06BBE-S Скачать руководство пользователя страница 106](http://html1.mh-extra.com/html/sartorius/factory-series-fc06bbe-s/factory-series-fc06bbe-s_operating-instructions-manual_1199734106.webp)
Printouts
Purpose
This function enables you to print out
weights, other measured values and IDs.
You can format the printout to meet
different requirements.
Features
Line format: you can configure a data
ID code with up to 6 characters at the
beginning of each of the values to be
printed
Weight ID:
You can configure an extra line for
identification of each weighed or
calculated value using the code
S ID
Print application parameters:
You can generate a printout of the val-
ues configured for initialization of
an application before printing the
measured results
ISO/GMP-compliant printout:
You can print out parameters relating
to the weighing conditions
Print animal weights:
Application-defined, automatic printout
of animal weights or of animal weights
plus calculated weights after averaging
Optimizing interfaces:
– Use the highest possible baud rate
– Turn off interfaces that are not in use
– Optimize the amount of data to be
transfered
Configuring Printout Formats
For a number of application programs, you
need to set initialization values. All values
upon initialization or only the main values
can be automatically printed as soon as
you have configured this in the Setup
menu:
Auto print upon
initialization
Weights and calculated values can be
printed as numeric values either with a
preceding data ID code (numeric value
with 22 characters) or without one
(numeric value only 16 characters). See
also the
section on
Line format
in the
chapter entitled “Data Output Functions”.
You can generate an ISO/GMP print-out
always or only for calibration/adjustment
or turn off this option. See also page 107.
Generating an ISO/GMP Printout
In the Setup menu, you have a choice
of three settings:
– No ISO/GMP printout generated (
Off
)
– ISO/GMP printout generated only for
calibration/adjustment (
Only for
calibration/adjustment
)
– Every printout is an ISO/GMP-compliant
report (
Always on
)
Auto print checkweighing results:
automatic printout of a weight when
it lies within the preset limits at stability
Auto print with time-controlled func-
tions: automatic printout of weights
after a preset time has elapsed or at
a defined time
Printout of intermediate or final evalua-
tion from the application 3 memory
(totalizing, formulation and statistics);
generate by pressing the
MR
soft key
Generating Printouts Acceptable for
Legal Metrology:
You can configure the Setup menu of
the scale to generate data records that
are acceptable for legal metrology on
a Sartorius printer:
– YDP02
– YDP03
– YDP01IS
– YDP01IS Label
– YDP02IS
– YDP02IS Label
– YDP04IS
– YDP04IS Label
106