Printout Parameters (Overview)
o factory setting
√
user-defined setting
Setup –
Application-
Stability parameter
Without stability
1)
Printout
defined
ο
With stability
Print on request,
ο
Off
then tare
On
Auto print upon
ο
Off
initialization
All values
(of the application)
Only main values
Configured printout
See separate chapter starting on the next page
FlexPrint
ο
Off
On
Automatic
Stability parameter
ο
Without stability
output of
At stability
displayed value
3)
Automatic after weight change
2)
Stop auto print
Use print key
p
ο
Not possible
Time-dependent
ο
1 display update
auto print
2 display updates
10 display updates
4)
100 display updates
4)
Output to
Serial communication
ο
Application-defined output
interface port
(PERIPHERALS)
Automatic output of displayed value
Line format
For raw data (16 characters)
ο
For other apps/GLP (22 characters)
ISO/GLP/GMP
ο
Off
printout
Only for calibration/adjustment
Always on
Identification
Lot ID (L ID):
Enter 20 characters max
ID1:
ID1
ID2:
ID2
ID3:
ID3
ID4:
ID4
Factory settings:
No
only printout
Yes
1
)
= When the scale is used in legal metrology, this setting may be configured for control purposes only; printout of values is not permitted
2
)
= Autoprint when load change is >10 d and stability is reached: no printout until residual difference in load value > 5 d
3
)
= To enable automatic output of the displayed value, the corresponding option (“Automatic output of displayed value”) must be activated
under the next menu item at this level, “Output to interface port”
4
)
= These settings may not be changed on verified scales
31