MP8 Interface Emulation
Purpose
With the MP8 interface emulation
function, you can connect peripheral
devices of the MP8 generation that
have separate AC power supplies (such
as the 73822... Data Control terminal,
a YFC..., or a YDI 50 Z Data Input
dedicated keyboard, for example) to
your scale.
Features
– The scale can be used only to determine
weights.
– The interface communicates exclusively
in the MP8 binary protocol.
– Select the application program and
the program index for MP8, as well
as individual application parameters,
in the Setup menu.
– See next page for menu setting options.
Preparation
– Activate the MP8 emulation mode*:
§
Press
H
§
Select the
factory settings
and confirm: press the
q
soft key
repeatedly and then the
O
soft key
§
Select
Reset to MP8
:
press
q
soft key and
O
soft key
§
Select
Yes
and press
l
to confirm
> The terminal is restarted
* follow the same procedure to return
from MP8 emulation back to factory
setting
Factory Settings
Each parameter category has a factory
setting. To restore the factory settings,
select this item in the Setup menu and
select
YES
to confirm.
The following parameters are not
restored to factory settings when you
activate this function:
– Language
– Password
– Display contrast
– Time (clock)
36