![Sartorius Factory Series FC06BBE-S Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/sartorius/factory-series-fc06bbe-s/factory-series-fc06bbe-s_operating-instructions-manual_1199734059.webp)
Preparation
To calculate a value in percent, the reference percentage must be known. This value can
be entered into the weighing-in-percent program in one of three ways:
– The last reference percentage entered is loaded and displayed when you turn on the
scale. Place the corresponding weight on the balance and initialize the weighing-in-
percent program.
– With automatic initialization switched on (Setup: App: Auto start application when power
goes on: On), the scale goes into the “weighing in percent” mode when you turn it on
and loads the last reference percentage entered as well as the corresponding reference
weight.
– Enter the reference weight using the numeric keys and store it (
Wxx%
soft key).
●
Turn on the scale: press
e
> Sartorius logo is displayed
●
Select the Weighing in percent application in the Setup menu: press
H
●
Select
Application parameters
: press the
q
soft key 2
+
,
then the
O
soft key once
●
Select
Application 1 (basic settings)
: press the
O
soft key
●
Select
Percent weigh.
: repeatedly press the
Q
or
q
soft key
●
Confirm
Percent weigh.
: press the
O
soft key
Percent weigh.
Weight storage
ο
Display accuracy
accuracy
+1 decimal place
+2 decimal places
Decimal places
None
for readout in pct.
1 decimal place
ο
2 decimal places
3 decimal places
4 decimal places
5 decimal places
6 decimal places
Display calculated
ο
Residue
value
Loss
Ratio1 (DR)
Ratio2 (OR)
ο
= factory setting
see also the “Application Parameters (Overview)” in the chapter entitled “Configuration”
●
Save settings and exit the Setup menu: press the
oo
soft key
Equations
The following equations are used for the corresponding calculations:
Residue (weighing in percent) = Current weight / 100% weight x 100%
Loss (percent-DIFF:)
= (Current weight – 100% weight) / 100% weight x 100%
Ratio1 (percent-Ratio 1:)
= (100% weight – current weight) / current weight x 100%
Ratio2 (percent Ratio 2:)
= 100% weight / current weight x 100%
Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring (not during alphanumeric input),
– printing,
you can also access the following
functions from this application:
Calibration/Adjustment
●
Press the
isoTST
soft key
> See “Calibration/Adjustment” for
further instructions
Toggling to the Next Application
●
Press
D
> See the section on the corresponding
application program for further
instructions
Setup (setting parameters)
●
Press
H
> See “Configuration” for
further instructions
Turning Off the Scale
●
Press
e
> The scale shuts off
59