20 Sartoclear
®
Pilot Filter Holder
Operation
t
Insert the filter cartridges one after the other, according to their labeling, on
the bottom guide rods . Pay attention to the alignment (up / down) of the filter
cartridges (see operating instructions for the filter cartridges) .
t
Push the filter cartridges towards each other until there is no longer any free
space between them .
t
Check the fit of the filter cartridge . The notches of the filter cartridge must be
positioned on the bottom guide rods of the filter holder .
t
Ensure that the filter cartridge is seated evenly and flush inside the filter holder .
t
Place the manifold plate (1) for the filtrate outlet according to its labeling onto
the bottom guide rods .
t
Push the manifold plate onto the filter cartridges .
t
Push the clamping plate (1) onto the manifold plate .
t
Position the holding clamps as closely as possible to the clamping plate (1)
(see Chapter "5 .1 Moving Holding Clamps", page 17) .
Move the Filter Cartridges Towards the Pressure Plate
Procedure
t
Move the inner holding clamps (1) on the two pull rods to the respective right-
hand position (see Chapter "5 .1 Moving Holding Clamps", page 17) .
t
Move the filter cartridges as close as possible to the pressure plate by pulling on
both pull rods (2) at the same time . The cylinder stroke should be kept as small as
possible .
t
Move the outer holding clamps (3) on the two pull rods to the right-hand
position .
1
1
1
1
3
1
2