- 20 -
F07:
<Purpose> To match a temp. sensor temperature with 1/2H air temperature
(Zero calibration of temp. sensor)
<Operation> Input service code in F06 prior to use this mode.
Input F07 and press SET key to display ‘00.0’ (Factory default).
Setting range is ‘-4.9’~’04.9’.
Change a value by pressing key and key.
Press SET key to memorize the value and to return to chamber
temperature display.
F08:
<Purpose> To calibrate a cascade sensor temperature
<Operation> Input service code in F06 prior to use this mode.
Input F08 and press SET key to display ‘00.0’ (Factory default).
Setting range is ‘-9.9’~’09.9’.
Change a value by pressing key and key.
Press SET key to memorize the value and to return to chamber
temperature display.
F12:
<Purpose> To display a temp. sensor temperature
<Operation> Input service code in F06 prior to use this mode.
Input F12 and press SET key to display alternately F12 and “XX.X”
(chamber temperature). Press SET key to return to chamber
temperature display. Decimal point is displayed in the 3 digits indication,
but minus “-“ is not displayed.
Ex) “-79.5
䉝
”
䊻㻌
Indicated as “79.5”
F13:
<Purpose> To display a cascade sensor temperature
<Operation> Input service code in F06 prior to use this mode.
Input F13 and press SET key to display alternately F13 and “XXX”
(present temperature in cascade sensor). Press SET key to return to
chamber temperature display.
F14:
<Purpose> To display a filter sensor temperature
<Operation> Input service code in F06 prior to use this mode.
Input F14 and press SET key to display alternately F14 and “XXX”
(present temperature in filter sensor). Press SET key to return to
chamber temperature display.
F15:
<Purpose> To display an AT sensor temperature
<Operation> Input service code in F06 prior to use this mode.
Input F15 and press SET key to display alternately F15 and “XXX”
(present temperature in AT sensor). Press SET key to return to
chamber temperature display.
F16:
<Purpose> Simply passing through if entered by mistake
<Operation> Press SET key in “F16” to return to chamber temperature display.
F17:
<Purpose> Display of model code and initialization of non-volatile memory
<Operation> Service code should be input in F06 prior to use this mode.
Input F17 and press SET key to display ‘00X’.
Change a value by pressing key.
Press SET key to memorize and return to chamber temperature display.
Model code ‘008’: MDF-U55V