- 23 -
F28:
<Purpose> Display of delay time to start measuring running ratio
(2hrs timer; 000~120 min)
<Operation> Service code should be input in F06 prior to use this mode.
Input F28 and press SET key to display alternately F28 with ‘xxx’
(present count value for delay time to start measuring running rate).
Press SET key to return to chamber temperature display.
When a delay time expires (a value reaches to ‘120’), unit will start
measuring running ratio.
F29:
<Purpose> Display of delay time to start diagnosing running ratio
(8hrs timer; 000~480 min)
<Operation> Service code should be input in F06 prior to use this mode.
Input F29 and press SET key to display alternately F29 with ‘xxx’
(present counting value for delay time to start diagnosing running ratio).
Press SET key to return to chamber temperature display.
8hours timer start counting after 2hours timer expires.
When a delay time expires (a value reaches to ‘480’), unit will start
diagnosing running ratio.
F30:
<Purpose> ROM version is displayed
<Operation> Service code should be input in F06 prior to use this mode.
Input F30 and press SET key to display alternately F30 with “X.XX”
(present ROM version).
Press SET key to return to chamber temperature display.
F31:
<Purpose> Setting of buzzer during filter alarm occurs
<Operation> Service code should be input in F06 prior to use this mode.
Input F31 and press SET key to display “001” (Factory default).
Change to alternative value ‘000’ or ‘001’ by pressing key.
Press SET key to revert to chamber temperature display.
000: Off
001: On
F32:
<Purpose> Display of accumulation time of condensing fan motor
<Operation> Service code should be input in F06 prior to use this mode.
Input F32 and press SET key to display alternately F32 with ‘xx.x’
(accumulation time).
Press SET key to return to chamber temperature display.
F44:
<Purpose> Adjustment for difference of power supply voltage
<Operation> Input F44 and press SET key to display ‘000’ (Factory default).
Setting range is ‘000’~’003’.
Press SET key to return to chamber temperature display.
Input value
Adjustment of display
000
0% (Power supply voltage is not increased)
001 3%
increased
002 5%
increased
003 7%
increased
F50:
<Purpose> Setting of alarm delay time
<Operation> Input F50 and press SET key to display ‘015’ (Factory default).
Setting range is ‘000’~’015’.
Change a value by pressing key and key.
Press SET key to memorize the value and to return to chamber
temperature display.