5-5 Device Value Sorting
1) Measuring ranges
L: 20.000 μH
~
2000 H (auto-range)
C: 200.00 pF
~
20.00 mF (auto-range)
R: 20.000 Ω
~
200.0 MΩ (auto-range)
2) Setting and Measuring procedure
①
Perform the OPEN/SHORT calibration.
For details, refer to the section of "OPEN/SHORT calibration".
(The OPEN/SHORT calibration is recommended for more
accurate measurement.)
②
Press
L/C/R AUTO/MANU
button to select an appropriate
manual measurement function.
This button allows you to select it as follows.
Auto LCR mode
⇒
Ls or Lp
⇒
Cs or Cp
⇒
Rs or Rp
⇒
DCR
⇒
Auto LCR mode
Note:
• The sorting mode is not available in the Auto LCR mode.
•
S/P
button allows you to select the series mode (Ls/Cs/Rs) or the
parallel mode (Lp/Cp/Rp).
③
Connect a device to the measuring terminals as a reference.
After the reading has stabilized, press the
SORT
button to launch
the sorting mode, and the reading will be saved as a reference.
[PASS] will be indicated as shown below.
Note:
If you press the
SORT
button while the meter is reading outside
limits (OL) or while the reading is less than 200 counts, the sorting
mode is not available.
④
To set the decimal point, range, or tolerance against the
reference value, press the
SETUP
button. The saved reference
value will be shown on the display.
5-4-3 Resistance (R) measurement
1) Measuring ranges
R: 20.000 Ω
~
200.0 MΩ (auto-range)
2) Measuring procedure
①
Press the
L/C/R AUTO/MANU
button to select Rs or Rp function.
②
Press the
TEST FREQ
button to select a measuring frequency.
③
Press the
S/P
button to select Rs or Rp.
Any sub parameter will not be shown.
④
Connect a DUT to the measuring terminals.
⑤
Read the display.
5
-4-4 DC Resistance measurement
1) Measuring ranges
R: 200.00 Ω
~
200.0 MΩ (auto-range)
2) Measuring procedure
①
Press the
L/C/R AUTO/MANU
button to select DCR function.
②
Connect a DUT to the measuring terminals.
③
Read the display.
Example