Value
Setting
Capacitance
Low
Parallel
High
Series
Inductance
Low
Series
High
Parallel
For details, refer to the section 4-7.
― 22 ―
― 21 ―
[5] MEASURING PROCEDURE
5-1 Connection of DUT (Device Under Test)
DUT's (Devices Under Test) may be connected to the meter as follows.
• Directly insert axial or radial component leads to the directly
connectable measuring terminals.
• Attach the clipping leads "CL-700" (bundled item).
Both Rs and Rp are part of the equivalent circuit of capacitors and
inductors.
To measure capacitance and inductance, refer to the settings as
shown in the table below.
4-15-3 Principle of OPEN/SHORT calibration
Z
M
is defined as total impedance measured to a DUT by a test fixture
which has some parasitic impedance.
Z
M
= (Rs + jωLs) + ((Go + jωCo)
-1
// Z
DUT
)
Z
DUT
is impedance of the DUT.
The OPEN/SHORT calibration reduces the effect of (Rs + jωLs) and
(Go + jωCo).
Rs
Z
M
Ls
Co
Go
Z
DUT