― 24 ―
― 23 ―
1. Do not use the instrument if the meter or the clipping leads look
damaged.
2. Make sure the clipping leads are not broken.
WARNING
• Make sure the battery condition is good after power-on.
Replace the battery with new one if the battery power is low.
• The directly connectable measuring terminals accept axial or
radial component leads up to 1.0 mm in diameter. Inserting
thicker leads may damage the measuring terminals.
CAUTION
Perform the OPEN/SHORT calibration before measurement to
ensure the safety and the accuracy.
For details, refer to the section 4-11.
5-2 Pre-operational Check
• Attach the SMD clipping leads "CL-700SMD"
(separately available accessory).
The guard line provides a shield for DUT, preventing interference
when measuring high-impedance devices.
CL-700SMD Electrical Characteristics
Parameter
Test Condition 100/120 Hz
1 kHz
10 kHz 100 kHz
Resistance: Rs
Clip opened
< 0.05 Ω < 0.10
Ω < 0.10 Ω < 0.10 Ω
Capacitance: Cp Clip closed
< 5.0 pF < 5.0 pF < 5.0 pF < 5.0 pF
Inductance: Ls
Clip closed
< 1.0 μH < 1.0 μH < 0.5 μH < 0.5 μH
• Temperature: 23
℃
±5
℃
, Humidity:
≦
80 % R.H.