5-4 Manual LCR Mode measurement
1. Do not apply any voltage or current to the measuring terminals.
2. Measuring live circuit may damage the meter.
3. Do not touch any metal part of the clipping leads nor the leads
of the DUT while measuring.
WARNING
• Discharge the capacitor before any measurement.
CAUTION
The meter will be in Auto LCR mode after power-on.
Pressing the
L/C/R AUTO/MANU
button switches to Manual LCR
mode and this button allows you to select the L/C/R as follows.
Auto LCR mode
⇒
Ls or Lp
⇒
Cs or Cp
⇒
Rs or Rp
⇒
DCR
⇒
Auto LCR mode
(Above each function except Auto LCR mode is in Manual mode.)
5-3 Auto LCR Mode Measurement
1. Do not apply any voltage or current to the measuring terminals.
2. Measuring live circuit may damage the meter.
3. Do not touch any metal part of the clipping leads nor the leads
of the DUT while measuring.
WARNING
• Discharge the capacitor before any measurement.
CAUTION
1) Measuring ranges
L: 20.000 μH
~
2000 H (auto-range)
C: 200.00 pF
~
20.00 mF (auto-range)
R: 20.000 Ω
~
200.0 MΩ (auto-range)
2) Measuring procedure
①
Press the POWER button to turn the meter on.
②
Press the
TEST FREQ
button to select a measuring frequency.
③
Connect a DUT to the measuring terminals.
④
Read the display.
Note:
D/Q/θ/ESR/Rp button is not available.