27
-
②
Memory data
Transfer the data stored in the memory in a block as a unit.
Transfer the measurement mode, upper limit value, lower limit value and number of the data at first and
followed by all data in blocks.
When the measured value exceeds the upper limit value, add the letter
「
H
」
in front of the measured
value, and letter
「
L
」
for the value under the lower limit value.
For the values except above mentioned, add symbol mark space "
△
" in front of the measured values.
◎
Output format (Example: TG-H mode)
Index data
C: Command
X: Figure 1 letter
△
: Space
STX
C Block No. Measurement mode
Upper limit value Lower limit value Data No. ETX
[02H]
1 : XXXX
: TG-H
: H XX.X
: L XX.X
: XXXX
[03H]
When H/L does not point, H
△
0.0 : L
△
0.0 is outputted.
Repeated data
STX
C Measured value
ETX
[02H]
1 :
△
XX.X
[03H]
When no data is in the applicable memory, NO DATA is outputted.
Index data
STX
C
Block No.
ETX
[02H]
1 : XXXX
: NO DATA
[03H]
Basic protocol
【
Moisture meter
】
【
PC
】
Data
ACK/NAK
(In case of parity error, transmit NAK)
Data
ACK/NAK
EOT
ACK