26
-
Communication output
①
Data while measuring
When the function of the communication with PC is ON, transfer the measured values
(values on hold) to PC.
Transmit the measurement mode, upper limit value, lower limit value, measured values with the first
transmission of the data.
When no upper and lower limit values are set, transfer the
「
0
」
.
When one of the mode, upper limit value or lower limit value, is changed on the way of the
measurement, transfer the measurement mode, upper limit value and lower limit value again as same
at the first time.
When the measured value exceeds the upper limit value, add the letter
「
H
」
in front of the measured
value, and letter
「
L
」
for the value under the lower limit value.
For the values except above mentioned, add symbol mark space "
△
" in front of the measured values.
◎
Output format (Example: TG-H mode)
First output
C: Command
X: Figure 1 letter
△
: Space
STX
C Measurement mode
Upper limit value Lower limit value Measured value
ETX
[02H]
$
:
TG-H
: H XX.X
: L XX.X
:
△
XX.X
[03H]
When H/L does not point, H
△
0.0 : L
△
0.0 is output.
Second and after output
STX
C
Measured value
ETX
[02H]
$
:
△
XX.X
[03H]
Basic protocol
【
Moisture meter
】
【
PC
】
Data
ACK/NAK
(In case of parity error, transmit NAK)