SanIjET Pure Pipeless Disinfection Protocol
IMPORTANT! TOTAL DISINFECTION REQUIRES TWO DISTINCT STEPS
Remove residue with
thorough and effective
cleaning.
1) PREPARATION & CLEANING
a. Drain the bathing well of all water.
b. Rotate the jet assembly to remove from jet case.
d. Disassemble the jet assemblies (3 pieces each) and place them in the sink. Use a non-
abrasive antibacterial cleaner to remove any visible residue - rinse with fresh water.
e. Use a non-abrasive antibacterial cleaner to clean the bathing vessel, jet cases, drain
stopper and overflow mechanism–rinse with fresh water.
2) DISINFECTION
a. Spray the surface area of the jet assemblies and bathing vessel until thoroughly
coated with an EPA–registered disinfectant.
b. Allow to set for the required disinfection contact period (refer to manufacturer’s
specifications) then rinse surfaces with fresh water.
3) REASSEMBLE
a. Dry the jet assemblies and bathing vessel.
b. Reassemble and replace jets by aligning the 11 o’clock position.
AdditionAl notes
•
SANIJET Pipeless™ Hydromassage system cleaning and disinfection takes approximately 5 minutes when using Dispatch® Hospital Cleaner
Disinfectant with Bleach.
•
The SANIJET Pipeless Jet System does not require a “fill and drain cycle” during disinfection, for improved water conservation.
•
For applications requiring quick cleaning between treatments, we recommend a second set of “clean” jet assemblies for quick replacement.
IMPOR
TANT
• Always refer to your state and local guidelines for cleaning requirements.
• neVeR Use abrasive cleaners on any surface area.
• After each use, disinfect the system with the recommended disinfection procedures for maximum sanitation.
• BeFoRe Use check the jet assembly is properly installed in the case motor assembly.
• BeFoRe Use make sure that water is filled to at least 1” above the jets for proper operation.
• do not attempt to activate the jet system without water present.
Disinfection: Apply an EPA registered
disinfectant, covering 100% of all surface
areas for the required contact time, to kill
bacteria, viruses and fungi.
1
2
3
Recommended cleAneR / disinFectAnt
SANIJET recommends the use of DISPATCH® Hospital Cleaner Disinfectant with Bleach - EPA Reg. No. 56392-7; EST. No. 62801-MI-001. Kills bacteria
and viruses in 1 minute and tuberculosis in 30 seconds. Available in 32 oz trigger spray and 64 oz refill bottle.