![Saluki SE1022D Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/saluki/se1022d/se1022d_user-manual_3698021080.webp)
Tel: 886. 909 602 109 Email: [email protected]
www.salukitec.com
80
Fig.62 The Configuration Area of Reference Signal
Table 7. Reference Signal Configuration
Reference Source
External
Internal
Int. Sweep
External Ref Trigger
TTL Rising Edge
Sine Zero Crossing
Int. Frequency
User is required to input manually, frequency range from
1mHz to 102kHz, minimum frequency resolution is 1mHz.
Phase (
°)
Configure the phase shift of two quadrature reference signals
of PSD algorithm. Phase accuracy is 0.01
°
. Input range from
-180
°
to +180
°
.
Auto Phase
Disable/Enable
Note that:
After finishing all the settings, click the button
to finish the configuration for input signal. All
settings can be changed simultaneously and you just need to click the button once.
After changing the settings, you need to click to finish this configuration. Otherwise, changes of settings will
not be saved.
2. Frequency Sweep Configuration
The configuration area of internal frequency sweep is shown in the red box in Fig.63. The details of configuration are
shown in Table 8.