Fig.32 <Offset & Expand> Submenu
<Offset> setting: It is set through keyboard. The adjustable range is -100% to +100%. The minimal step is 0.01% and
0.00% is default.
<Expand> setting: It is set through keyboard. The adjustable range is 1 - 256 and 1 is default. If the setting value of
Expand causes the calculation result over 10V, the output will be seriously affected.
The settings of <Offset> and <Expand> will not affect the data display in dynamic area.
3.6.3 < Speed >: Setting the Output Speed
<Slow>
the data output speed of channel CH1 & CH2 is 10 Hz
<Fast>
the data output speed of channel CH1 & CH2 is 312.5 kHz
3.7
[SAMPLE]
[SAMPLE] submenu is used to control the sampling and storage. There are 5 settings in this submenu, including <Step
Time>, <Length>, <Select>, <Trigger Mode> and <Sample Mode>, as is shown in Fig.33. Soft key [CH A/B SWITCH]
is provided for switching between two channels. Users can also switch between two channels through knob.
The setting of [SAMPLE] menu should cooperate with the keys in CONTROL area. In this case, [SAMPLE] can control
the sample to start, pause and reset.