indicates the step frequency(unit:Hz).The SSLLD? 1 command will return the step
frequency of channel A. The range of f is 0-102kHz and the minimum resolution is 1mHz.
SSLGD(?) i {,x}
The SSLGD command sets or queries the percentage in <Log> internal sweep mode. The
parameter i represents the selected channel (A or B). The parameter x indicates the
percentage(unit:%). The SSLGD? 1 command will return the percentage of channel A in
<Log> frequency sweep mode. The range of x is 0-100 and the minimum resolution is
0.001. x=50 means that the percentage is 50%.
STLMD(?)i {,j}
The STLMD command sets or queries the step time in internal frequency sweep mode. The
parameter i represents the selected channel (A or B). The parameter j indicates the step time
(unit:ms).The STLMD? 1 command will return the step time of internal sweep mode. The
range of j is 1ms-100,000ms and the minimum resolution is 1ms.
SWRMD(?)i {,j}
The SWRMD command sets or queries the running mode in internal frequency sweep
mode. The parameter i represents the selected channel (A or B). The parameter j indicates
the running mode. The parameter j selects <STOP> (j=0) , <Single> (j=1) or <Loop> j=2).
HARMD(?) i,j {,k}
The HARMD command sets or queries harmonic detection. Parameter i and j are necessary
for transmitting this command. The parameter i represents the selected channel (A or B).
The parameter j selects harmonic 1(j=1) or harmonic 2 (j=2). The parameter k is an integer
from 1 to 32767. The HARMD i,j,k command will set the SE1022D to detect at the the
harmonic of reference frequency. The value of k is limited by k*f≤102kHz. If the frequency
of the harmonic exceeds102kHz, then j will be set automatically to the maximum value that
meets this condition.
SLVLD(?)i {,x}
The SLVLD command sets or queries the amplitude of the output synchronous sine wave.
The parameter i represents the selected channel (A or B). The parameter x determines the
amplitude of voltage (real number, unit: V). The range of x is 0.000Vrms≤x≤5.000Vrms
and x is rounded to 0.001Vrms.
SWVTD(?)i {,j}
The SWVTD command sets or queries the output type of <Sine Output>. The parameter i
represents the selected channel (A or B). The parameter j selects fixed amplitude output
(j=0), linear amplitude sweeping output (j=1) or logarithm amplitude sweeping output (j=2).
SVLLD(?)i {,x}
The SVLLD command sets or queries <Start> amplitude in <Sine Output> sweep mode.
The parameter i represents the selected channel (A or B). The parameter x indicates the start
amplitude (unit:V).The SVLLD? 1 command will return the start amplitude of sweeping
amplitude of channel
A. The range of x is 0.001Vrms≤x≤5.000Vrms and the minimum resolution is 0.001Vrms.
SVULD(?)i {,x}
The SVULD command sets or queries <Stop> cutoff amplitude in <Sine Output> sweep
mode. The parameter i represents the selected channel (A or B). The parameter x indicates
the cutoff amplitude (unit:V). The SVULD? 1 command will return the cutoff amplitude of
sweeping amplitude of channel A. The range of x is 0.000Vrms≤x≤5.000Vrms and the