To prevent or reduce the possible damage to the internal hardware of the spectrum analyzer
due to the mutual interference caused by several sets of equipment through the power
supply, especially the peak pulse interference caused by high-power equipment, it is best to
use a 220V/110V AC stabilized-voltage power supply to supply power for the spectrum
analyzer.
Bad grounding or failure of the power supply may cause damage to the instrument, even
personal injury. Before connecting the power supply of spectrum analyzer, make sure
ground wires of the instrument and its power supply are in good contact.
Use a power socket with protective grounding and do not substitute the protective ground
wire with an external cable.
3.1.4
Electrostatic Protection
ESD can damage or destroy electronic components and equipment. Antistatic measures are usually taken to prevent ESD
damage: conductive table–mat and wrist–strap combination as well as conductive floor–mat and heel–strap combination. Both
types, when used together, provide a significant level of ESD protection. Among the two, only the table–mat and wrist–strap
combination provides adequate ESD protection when used alone. To ensure user safety, the static–safe accessories must
provide at least 1 MΩ resistance to ground.
Please use the following antistatic measures correctly to reduce the damage of electrostatics:
Ensure all the instruments are grounded correctly to prevent the electrostatics being generated.
Before touch the joint and core wire or do any assembly operation, do wear antistatic wrists or take other antistatic measures.
The above antistatic measures shall not be taken where the voltage is larger than 500V.
3.1.5
First Power Up
1)
Connect the provided power cord instrument end to the power socket on rear panel.
2)
Ensure the AC power is compliant to requirement.
3)
Connect the other end to AC power
4)
Press the power button on front panel.
5)
Screen lighted and it will take around 30 sec to run the self-test and adjust. Waiting for the process to be done.
6)
For an accurate measurement, please warm up the spectrum analyzer for 30min before measurement.
Содержание S3531 series
Страница 1: ...S3531 Series Spectrum Analyzer User Manual Saluki Technology Inc...
Страница 32: ...Tel 886 2 2175 2930 Email sales salukitec com www salukitec com 20 Fig 5 15 TOI Distortion Measurement...
Страница 46: ...Tel 886 2 2175 2930 Email sales salukitec com www salukitec com 34 6 11 Meas...
Страница 71: ...Tel 886 2 2175 2930 Email sales salukitec com www salukitec com 59 9 Appendix A Programming Manual...