Tel: 886.2.2175 2930 Email: [email protected]
www.salukitec.com
14
Figure
5
-8: 3dB Bandwidth Test Result
5.5
Adjacent Channel Power
Following example uses a S1103 signal generator and generates a 500MHz, -10dBm continuous signal.
Example
1)
Connect the signal generator output port to S3531 RF test port.
2)
Press
【
Preset
】
key to return the instrument to default status
3)
Set test frequency.
Press
【
FREQ
】
key
Select [Center FREQ] and input 500MHz
4)
Press
【
Meas
】
key. Select [Adjacent Power]
The trace will be displayed as shown in the figure below.
5)
Read out the result