Self Calibration: The self calibration routine can optimize the precision of the oscilloscope to fit the ambient
temperature. To maximize the precision, you should perform the self calibration once the ambient temperature
changes by 5
℃
or more. Follow the instructions on the screen.
Tip: Press any menu button on the front panel to remove the status display and enter a corresponding menu.
5.5.5
Cursor measurement
Push the CURSOR button to display the Cursor Menu.
Options
Settings
Comments
Type
Off
Voltage
Time
Track
Select a measurement cursor and display it.
Voltage measures amplitude while Time measures time and frequency.
Source
CH1
CH2
CH3
CH4
MATH
REFA
REFB
Select a waveform to take the cursor measurement.
Use the readouts to show the measurement.
Select Cursor
S
E
S indicates Cursor 1. E indicates Cursor 2.
A selected cursor is highlighted, which can be moved freely. Both cursors can be selected and moved
at the same time. The box behind the cursor displays the location of the cursor.
Delta
Display the
difference (delta)
between the
cursors.
Display the measurement in the box under this option.
Moving Cursors: Press the key near Select Cursor to select a cursor and turn V0 to move it. Cursors can be moved
only when the Cursor Menu is displayed.