41
Options
Settings
Comments
27
LFF
The time between the first falling edge of source 1 and the last falling edge of source 2.
28
LFR
The time between the first falling edge of source 1 and the last rising edge of source 2.
29
LRF
The time between the first rising edge of source 1 and the last falling edge of source 2.
30
LRR
The time between the first rising edge of source 1 and the last rising edge of source 2.
31
FFR
The time between the first falling edge of source 1 and the first rising edge of source 2.
32
FRF
The time between the first rising edge of source 1 and the first falling edge of source 2.
Off
Do not take any measurement.
Use the knob V0 or the functional keys F3, F4 to select the type of measurement.
Taking Measurements: For a single waveform (or a waveform divided among multiple waveforms), up to 8
automatic measurements can be displayed at a time. The waveform channel must stay in an ‘ON’ (displayed) state to
facilitate the measurement. The automatic measurement can not be performed on reference or math waveforms, or
in XY or Scan mode.
5.5.3
Acquisition System
Push the ACQUIRE button to set the acquisition parameter.
Options
Settings
Comments
Type
Real Time
Equ-Time
Acquire waveforms by real-time digital technique.
Rebuild waveforms by equivalent sample technique.