Tel: 886.2.2175 2930 Email: [email protected]
11
3
Getting Started Guide
3.1
Installation
To keep proper ventilation of the oscilloscope in operation, leave a space of more than 5cm away from the top and
the two sides of the product.
3.1.1
Power Supply
Use a power supply that delivers 90 to 240 V
RMS
, 45 to 440 Hz.
3.1.2
Power Cord
Use only power cords designed for this product.
3.2
Functional Check
Follow the steps below to perform a quick functional check to your oscilloscope.
3.2.1
Power on the oscilloscope
Plug in the oscilloscope and press the ON/OFF button. The default Probe option attenuation setting is 10X.
3.2.2
Connect the oscilloscope
Set the switch on the probe to 10X and connect the probe to Channel 1 on the oscilloscope. First, align the slot in the
probe connector with the protuberance on the CH1 BNC and push to connect; then, turn to right to lock the probe in
place; after that, connect the probe tip and reference lead to the PROBE COMP connectors. There is a mark on the
panel: Probe COMP ~2V@1KHz.
3.2.3
Observe the waveform
Press the AUTOSET button and you should see within a few seconds a square wave of about 2V peak-to-peak at
1kHz in the display. Press the CH1 MENU button twice to remove Channel 1. Push the CH2 MENU button and
repeat Step 2 and Step 3 to observe Channel 2.