6.
The coupling method is only applicable to the analog signal.
7.
Turn the timebase knob “SEC/DIV” and adjust the time until you get a stable display of waveforms. See the
picture below.
Pulse Width Trigger
To observe digital signal by using the pulse width trigger, please follow the steps as below:
1.
Push the TRIG MENU button to see the trigger menu.
2.
Push F1 button to select “Pulse” as the trigger type.
3.
Push F2 button to select trigger source (D0~D15).
4.
Push F3 button to select the pulse polarity ( positive or negative).
5.
Push F4 button to select the trigger mode as Auto or Normal.
6.
The Coupling method is only applicable to the analog signal.
7.
Push the ‘Next Page’ button and push F4 button to select the trigger event (=,≠
,
>
,
<).
8.
Push F5 button to select the the pulse width of the trigger, and turn VO knob to adjust the trigger pulse width.
9.
Turn the timebase knob “SEC/DIV“ and adjust the time until you get a stable display of waveforms. See the
picture below.