12
3.2.3
Observe the waveform
Press the AUTOSET button and you should see within a few seconds a square wave of about 5V peak-to-peak at
1kHz in the display. Press the CH1 MENU button twice to remove Channel 1. Push the CH2 MENU button and
repeat Step 2 and Step 3 to observe Channel 2.
3.3
Probe Examination
3.3.1
Safety
When using the probe, keep your fingers behind the guard on the probe body to avoid electric shock. Do not touch
metallic portions of the probe head while it is connected to a voltage source. Connect the probe to the oscilloscope
and connect the ground terminal to ground before you start any measurements.
3.3.2
Use of Probe Check Wizard
Every time you connect a probe to an input channel, you should use the probe check wizard to verify that this probe
is operating correctly. There are two ways to do this:
1)
Use the vertical menu (for example, push the CH1 MENU button) to set the Probe option attenuation factor.
2)
Press the PROBE CHECK button to use the Probe Check Wizard and configure the probe option attenuation
factor properly following menu prompts.