19
I
GB
Table 4
Recommended separation distances between portable and mobile RF communications
equipment and the device.
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmittters) and the device as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 3.5 √
—
P
80 MHz to 800 MHz
d = 1.2 √
—
P
800 MHz to 2.7 GHz
d = 2.3 √
—
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres
(m) can be determined using the equation applicable to the frequency of the transmitter, where
P
is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
2 YEAR GUARANTEE
FKA Brands Ltd guarantees this product from defect in material and workmanship for a period of 2 years
from the date of purchase, except as noted below. This FKA Brands Ltd product guarantee does not cover
damage caused by misuse or abuse; accident; the attachment of any unauthorised accessory; alteration
to the product; or any other conditions whatsoever that are beyond the control of FKA Brands Ltd. This
guarantee is effective only if the product is purchased and operated in the UK / EU. A product that requires
modification or adaptation to enable it to operate in any country other than the country for which it was
designed, manufactured, approved and / or authorised, or repair of products damaged by these modifications
is not covered under this guarantee. FKA Brands Ltd shall not be responsible for any type of incidental,
consequential or special damages.
To obtain guarantee service on your product, return the product post-paid to your local service centre along
with your dated sales receipt (as proof of purchase). Upon receipt, FKA Brands Ltd will repair or replace, as
appropriate, your product and return it to you, post-paid. Guarantee is solely through Salter Service Centre.
Service of this product by anyone other than the Salter Service Centre voids the guarantee.
This guarantee does not affect your statutory rights.
For your local Salter Service Centre, go to
www.salterhousewares.co.uk/servicecentres
Содержание BPA-9201
Страница 1: ...AUTOMATIC ARM BLOOD PRESSURE MONITOR Instructions and Guarantee BPA 9201 EU ...
Страница 5: ...5 I GB LCD display signal ...