Scale calibration  

Calibration Configuration Settings –  

The scale is configured from the factory to match the specified settings for each unit, as defined by the product 
specifications and sales brochure.  Modification of the setting can be accomplished by altering user configuration 
settings P7-P10. 

Caution:

 Calibration and/or configuration of calibration settings of your scale should be 

accomplished by a trained service technician using certified weights to ensure proper operation and accuracy.  
Calibration is not covered under warranty.  

 
Setup Controls 

 

Hold

Print

Tare

       

       

kg

 

 
Saves data and move down 
to next Parameter setting 
 .

 

On/Off

Zero

 

 
 
 
 

Exits setup mode

 

 

 
 
 

 

kg

 

     

Moves flashing digit

 

Units

 

 

  

 

 

Change flashed digits

 

 
 
 

 
 

Config 

Menu 

Avail. 

Settings 

Default Definition 

Detailed 

Setting 

P7 00-31 

See table 2A 

Displayed Resolution

(00) = 500       (08) = 2400     (15) = 7000     
(01 )= 600       (09) = 2500     (16) = 7500 
(02 )= 750       (10 )= 3000     (17) = 8000 
(03 )= 800       (11) = 3500     (18) = 10,000 
(04) =1000      (12 )= 4000     (19) = 12,000 
(05 )=1200      (13) = 5000     (20) = 15,000 
(06) =1500      (14) = 6000    (21-31) = N/A 
(07) = 2000         

P8 0,1,2 

See table 2a 

Division Increment 

0 = 1 ; 1 = 2 ; 2 = 5 

P9 0-5 

See table 2a 

Decimal Position 

0 = 123456 ; 1 = 12345.6; 2 = 1234.56 ;  
3 = 123.456 ; 4 = 12.3456 ; 5 = 1.23456  

P10 0, 

See table 2a 

Calibration weight  

0 = KG ; 1 = LB 

 

 
Table 2a 
 

You may choose to configure your scale for a higher resolution. The factory does not recommend increasing the 
resolution above 7,500 divisions for a stable weight reading.  Certain environments may cause the scale to be 
unstable at factory settings, reduce the # of division settings to increase your stability.   

Calibration Menu  

 
 
 
 
 
 

 
 
 
 
   

Calibration can be done with 25% to 100% of requested load and can be calibrated with 1 or 2 calibration points 

1.

 

On/Off

Zero

+

Tare

   

 

 

 

 

 

  

 

Units

 

 

  

 

Hold

Print

 

kg

 

     

Moves flashing d

Tare

       

On/Off

Zero

 

       

kg

 

 
 
 

 

Exits setup mode 

 

 

igit

 

Saves data and move down 
to next Parameter setting 

Change flashed digits

 .

 

Содержание SBI-100

Страница 1: ...Model SBI 100 Operation Manual Manual xxxxxx Issue xx...

Страница 2: ...DOUBT SEEK ADVICE FROM A QUALIFIED ELECTRICIAN Routine maintenance To avoid the possibility of electric shock or damage to the machine always switch off the machine and isolate from the power supply...

Страница 3: ...pecifications 15 Error Messages 15 Declarations of Compliance 16 Contact Information Technical Support 16 Product Identification Table Part Number Capacity Region Power Supply s 816965001620 75 kg x 0...

Страница 4: ...t Power Adapter Socket Power Supply Connecting the Indicator to the Platform 6VDC 500mA 230V 50Hz Wire the cable attached to the base as shown if necessary 4 2 3 1 1 RED 2 RED 3 RED 4 RED 1 Red Excita...

Страница 5: ...0 Weight is zero On Off Zero On Off Zero kg lb Unit of measure Flashing LED means that the weight reading is not stable Units kg lb Unit of measurement NET Tare is activated Hold Print Hold Print CH B...

Страница 6: ...tion Turning On and Zeroing the Scale On Off Zero ZERO 0 Turning off the scale On Off Zero Selecting Unit of Measure Units kg lb kg lb Using the Tare 1 2 Tare NET 3 NET 4 NET Removing the Tare On Off...

Страница 7: ...old function with manual release This function needs to be setup in Parameter P3 1 shown on page 10 1 Hold Print Hold 2 3 Removing Hold Hold Print Hold Print Function For communications to a printer o...

Страница 8: ...s Hold Print Entering setup On Off Zero Units Selecting parameter Tare Changing data within the parameter Units Saving data Tare Exiting setup On Off Zero ZERO 0 kg Moves flashing digit Units Change f...

Страница 9: ...ctable hold window from 3 to 50 divisions Will hold display reading once stable within a selectable weight range to release the hold button must be represses P3 0 P3 1 P3 2 P3 3 to 50 P4 x RS232 Seria...

Страница 10: ...tare LF H1H2H3 CR ETX U CR Changes the units LF u1u2 CR LF H1H2H3 CR ETX L CR Activates the hold function LF H1H2H3 CR ETX X CR Switches off the scale Indicator switches off Unrecognised command LF CR...

Страница 11: ...mple Gross xxxxx0 18lb Tare xxxxxx0 18lb Net xxxxxxx0 18lb Format LF Gross reading minus decimal point unit CR EXT LF Tare reading decimal point unit CR EXT LF Net reading minus decimal point unit CR...

Страница 12: ...olution 00 500 08 2400 15 7000 01 600 09 2500 16 7500 02 750 10 3000 17 8000 03 800 11 3500 18 10 000 04 1000 12 4000 19 12 000 05 1200 13 5000 20 15 000 06 1500 14 6000 21 31 N A 07 2000 P8 0 1 2 See...

Страница 13: ...0 kg lb 5 Single point calibration enter the same weight in again and move to number 7 For 2 point calibration enter in the second calibration weight between 25 100 full capacity Tare 6 7 Tare 8 9 Tar...

Страница 14: ...and LB P12 0 7 7 Power On zero range full scale 0 1 1 2 2 5 3 10 4 20 5 50 6 100 7 no limitation P13 00 15 3 Zero button range Full scale 00 1 06 20 12 5 01 2 07 no limit 13 10 02 3 08 1 14 20 03 4 09...

Страница 15: ...ill be turned off in two minutes automatically AC Adapter 10 5 12VDC 600mA with central positive 5 7 VDC 9 VDC Error Messages Error Message Definition Required Solution Troubleshooting 0 Weight above...

Страница 16: ...ications Le pr sent appareil num rique n met pas de bruits radio lectriques d passant les limites applicables aux appareils num riques de la Classe A prescrites dans le R glement sur le brouillage rad...

Отзывы: