NAME/CODE

DESCRIPTION

CODE/VALUE

F1

Graduations

Specifies  number  of  full-scale  graduations.  Value  should  be

consistent  with  legal  requirements  and  environmental  limits  on  the

useful system resolution.

500        1,000

1,500     2,000

2,500     3,000

4,000     5,000

6,000     8,000

10,000

√√√√

   12,000

20,000   30,000

40,000    50,000

F2

Span Gain

Span  Gain  is  related  to  A/D  integration  time.  The  larger  the  span

gain,  the  higher  the  internal  resolution,  but  the  slower  the  update

speed.  Note  that  the  scale  must  be  re-calibrated  whenever  this

parameter is altered. See Appendix C for more information.

25       50

75     

100

√√√√

150     200

F3

Zero Track

Band

Selects the range within which the scale will automatically zero. Note

that the scale must be in standstill to automatically zero. Selections

are in Display Divisions.

0d

0.5d

√√√√

1d

3d

5d

F4

Zero Range

Selects the range within which the  scale  may  be  zeroed.  Note  that

the indicator must be in standstill to zero the scale.

100%

√√√√

1.9%

F5

Motion Band

Sets the level at which motion is detected by comparing the present

display  update  with  the  previous  one.  If  motion  is  not  detected  for

two seconds or more, scale is in standstill and can process a Print or

Zero  command.  Maximum  value  varies  depending  on  local

regulations.

1d

√√√√

3d

5d

10d

F6

Digital Filter

Averages weight readings to produce higher stability. The higher the

filter  number,  the  greater  the  stability  but  the  slower  the  response

time. Choose 8 or 12 unless a very fast response is needed. Use the

“Auto” setting to automatically determine the filter length as you are

weighing.

1          2

4          8

12        16

24

F7

Overload Limit

Selects the desired formula which determines the point at which the

indicator  shows  overload.  All  selections  are  based  on  the  primary

unit selected in F8.

"FS" = Full scale in primary units.

FS

FS + 2%

√√√√

FS + 1d

FS + 9d

F8

Calib. Unit

Selects the primary base unit to be used in the calibration process.

Also the default unit for normal operation.

"1" = primary unit is lb.                       "2" = primary unit is in kg.

1

√√√√

2

Содержание PR-3000 Series

Страница 1: ...ruary 13 2004 Contents subject to change without notice Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont MN 56031 Tel 800 637 0529 Tel 507 238 8702 Fax 507 238 8271 E mail sales salter...

Страница 2: ...n Item 7 3 3 2 Taring an Item 7 8 Chapter 4 Advanced Features and Operation 9 4 1 Piece Counting 9 Chapter 5 Scale Configuration 10 5 1 Scale Configuration Overview 10 11 5 2 User A Menu 12 5 2 1 Ente...

Страница 3: ...internal battery 20 D 4 How long to charge the internal battery 20 LIST OF FIGURES 1 1 PALLET RUNNER 3000 Front Panel 4 3 1 PALLET RUNNER 3000 Display Detail 6 3 2 Function Keys Layout 7 5 1 PALLET RU...

Страница 4: ...ed after unpacking the scale A small amount of preparation is necessary before the scale can be used You will find this information in Chapter 2 Please read these instructions carefully and completely...

Страница 5: ...Engage the lift mechanism by lowering the hand control It may be necessary to hit the release button before doing so 7 Lift the pallet by cranking the handle in an UP DOWN direction a minimum of 3 com...

Страница 6: ...ING 0 Better known as the Center of Zero annunciator this light is active whenever the displayed weight is within 0 25 divisions of true zero N Indicates that the scale is displaying net weight G Indi...

Страница 7: ...cale is not at or below Gross zero 2 The scale is not in motion 3 The scale is not in overload see Appendix C for error codes ON Print This key is used to turn the scale ON The Print function is not a...

Страница 8: ...he display shows zero weight and turns the NET annunciator on 5 Place the material to be weighed on the container or pallet raise it off the floor and allow the weight indication to stabilize 6 Read t...

Страница 9: ...ontainer or pallet you must first tare the container or pallet off by pressing the TARE key 3 If the screen does not show 10 0 press the TARE key once The scale is prompting you to place ten identical...

Страница 10: ...the switch in the down position as shown at right 4 Power on the indicator by pressing the ON key The indicator shows F 1 to indicate that you are in Setup Menu mode CANNED A D CONVERTER SETUP SWITCH...

Страница 11: ...ound in Table 5 1 Factory set defaults are shown in bold with a checkmark 5 2 4 EXITING THE USER MENU 1 Power off the indicator by pressing and holding the down ZERO OFF key for 3 seconds 2 Move the S...

Страница 12: ...vel begins the sequence 0 199999 123456 A6 No of line feeds Actuates the function that allows entry of the desired number of line feeds to be printed in the Print Ticket Mode Valid only when A3 is set...

Страница 13: ...e indicator must be in standstill to zero the scale 100 1 9 F5 Motion Band Sets the level at which motion is detected by comparing the present display update with the previous one If motion is not det...

Страница 14: ...Places indicator into the zero calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F17 Span Calibration Places indicator into the span...

Страница 15: ...15...

Страница 16: ...e the zero point value The display will show EndC0 momentarily then revert back up to F16 At this time proceed to the F17 span calibration to complete indicator calibration 6 3 SPAN CALIBRATION F17 1...

Страница 17: ...Change the calibration test weight or check the input data Err1 The calibration test weight or the adjusted keyed in weight is smaller than 1 of the full capacity of the scale Change the calibration t...

Страница 18: ...Scale Capacity OPERATING TEMPERATURE RANGE 14 F to 104 F 10 C to 40 C POWER SOURCE 2 internal 6V lead acid batteries AC Adapter 12VDC 800 mA Also acts as charger CHARGER See Power Source above Charge...

Страница 19: ...eck all load connections Use F16 mode to view internal counts See Appendix C for more information Err 3 All Modes Non volatile memory read error One or more setup parameters have been lost Err 4 All M...

Страница 20: ...scale may be used for an additional 15 to 30 minutes without damage to the internal battery If the scale is used for more than 30 minutes after a low battery indication permanent damage may occur to t...

Отзывы: