FUNCTION SETTINGS
1.
Press [ON/OFF] to turn on the scale, the display will show flash digits and
then --------,
press
[PRINT]
, the display will show XXXX BAU value ( press
[UNIT]
to select among 1200,2400,4800,9600), our default setting is 9600.
Press
[ZERO]
to enter Parity mode, the display will show none
PAR
,press [UNIT] to
select odd,even,none (odd stands for 7 data bits with odd parity, even stands for 7
data bits with even parity,none stands for 8 data bit without parity)our default setting
is none.
Press
[ZERO]
to enter Print mode, the display will show X mod, press [UNIT]to
select 0,1,2,3 (0 stands for serial data output disabled,1 stands for demand mode by
press
[PRINT],2
stands for automatic print when scale is stable, 3 stands for continuous
print, our default setting is 1.press [ZERO] again to return to weighing mode.
2. Press [ON/OFF] to turn on the scale, the display will show flash digits and
then ----------,
press
[UNIT]
to enter auto off function, the display will show OFF
PWR
, press
[UNIT]
to select among 3,5,10,off ( 3 stands for 3minutes, 5 stands for
5minutes, 10 stands for10 minutes, off stands for inactivate auto off function )
press
[ZERO]
to select backlight, the display will show X BL ,press
[UNIT] to select
X
value (
X
stands for 0,1,2) , 0 stands for inactivate backlight function, 1 stands for
the backlight is always make function. 2 satnds for auto backlight .
Press
[ZERO]
to enter unit selecting function, the display will show ON X ( X stands
for the units
:
g,oz, lb,twt…tola) press
[UNIT]
to select ON (activate the selected
unit) or OFF ( inactivate the selected unit), press [ZERO] again to return to weighing
mode.
3. Switch the power to
“
ON
”
and than press [ON/OFF]to turn on the scale
。
The
display shows F-E-9 to 0 and than
“
- - - - - - -
”
,
press[PCS]key to enter ZERO track
function setting
,
display shows x.xd ZER
,
press [UNIT] key to select ZERO track
value
(
0.0d-0.5d-1.0d-1.5d-2.0d-3.0d
)
; press [ZERO]key to enter ZERO setting
,
display shows x.xd RTZ, press [UNIT] key to select ZERO value
(
0.5d-1.0d-2.0d-
3.0d
)
; press [ZERO] key to enter single segment setting or double segment setting ,
press [UNIT] key to switch between them ,SIN dsp means single segment setting
(default setting),db 1 dsp means double segment setting; press [ZERO] key to enter
FIL setting, press [UNIT] key to select filter rangn amond (0-1-2-3-4),”2”is default
setting; press [ZERO] key to select zero range when turn on the scale, press [UNIT]
key to select percent (2 PZR-5 PZR -10 PZR -20 PZR); press [ZERO] key to set zero
range, press [UNIT] key to select percent (5 KZR or10 KZR); press [ZERO] key to
set TARE range, press [UNIT] key select percent (50 KTR or100 KTR); press [ZERO]
key to enter overload protection setting, press [UNIT] key to select overload
protection value (9d OVR-1 OVR-2 OVR-10 OVR); press [ZERO] key to set beep
function, press [UNIT] key to choose turn on or turn off the beep, “ON”means
activate the beep, “OFF” means inactivate the beep; press [ZERO] key to unite the
function of TARE and ZREO, press [UNIT] to switch between them.“ON”means
unite function activate, “OFF”means unite function inactivate. press [ZERO]key
again to return weighing mode..
4. Press
[ZERO]
to return to weighing mode
.
COMPACT SCALE
/
BALANCE
INSTRUCTION MANUAL
Manual_FR-H_V1 ENG