Setup Menu Mode / 

Continued

To place the unit in Setup menu mode:

1.  Turn the power off to the unit.
2.  Locate  the  calibration  switch  on

the scale’s back panel and toggle

to the opposite position.

3.  While  holding  down  both  the

ZERO

 and 

PRINT

 keys, turn the

power back on.

4.  When  the  display  shows  "

F1

",

the unit  is  in  Setup  Menu  mode,

and  you  can  release  the  two

keys.  Shown  at  right  are  the  di-

rectional  and  SET  key  assign-

ments

SET

SETUP MENU CHART

F1

Display Mode

F6

Zero Range

F9

Calib. Unit.

0

1

F11

Display Check

F12

Calibration

Press ZERO

key to begin

Press ZERO

key to begin

0

1

100% 1.9%

To place the unit back into the Normal Operating mode:

1. With the scale on or off, toggle the Calibration Switch back to its original position. Turn the scale

back on without pressing any keys.

2. The display will go through a digit check, then settle into Normal Operating mode. All front panel

keys will now return to their normal mode of operation.

Содержание PC 150

Страница 1: ...on 1 2 September 14 2000 Contents subject to change without notice Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont MN 56031 Tel 800 637 0529 Tel 507 238 8702 Fax 507 238 8271 E mail s...

Страница 2: ...l communi cation port The scale can be configured to display its divisions in two separate modes non NTEP mode and NTEP mode The scale is shipped from the factory configured in non NTEP mode The NTEP...

Страница 3: ...The scale is now ready for use ON OFF SWITCH DC POWER JACK RS 232 PORT ID TAG CALIBRATION SWITCH COVER PC 150 Back Panel The female 9 Pin RS 232 connector is used to interface to a PC Shown at right a...

Страница 4: ...the displayed weight is within 0 25 divisions of true zero NET Indicates that the scale is displaying net weight GROSS Indicates that the scale is displaying gross weight TARE Indicates that the tare...

Страница 5: ...S Toggles between Gross and Net weight display if a tare has been established TARE Used to establish the weight on the platform as the tare weight You cannot establish zero or negative gross weight as...

Страница 6: ...off to the unit 2 While holding down the lb kg key turn the power back on 3 When the display shows A1 the unit is in User Menu mode and you can release the lb kg key Shown at right are the direc tion...

Страница 7: ...Serial Data Format Selects the data format to be transmitted via the serial port to a printer or computer 0 Consolidated Controls Format 1 Toledo 8213 Format 2 NCI 3825Format 3 Transcell Technology F...

Страница 8: ...Shown at right are the di rectional and SET key assign ments SET SETUP MENU CHART F1 Display Mode F6 Zero Range F9 Calib Unit 0 1 F11 Display Check F12 Calibration Press ZERO key to begin Press ZERO...

Страница 9: ...one level begins the test sequence Press ZERO key to begin sequence F12 Calibration Places scale into the calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZER...

Страница 10: ...y pressing the PRINT key or the TARE key 5 After setting the exact value press the NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then freez...

Страница 11: ...0 6 7 Segment Green LED OVER CAPACITY ANNUNCIATION 102 of Full Scale Capacity OPERATING TEMPERATURE RANGE 32 F to 104 F 0 C to 40 C PLATTER SIZE 13 x 13 330 mm x 330 mm POWER SOURCE AC Adapter 12VDC 5...

Страница 12: ...al Select the proper settings for the following using the scale s settings Baud Rate Data Bits Parity Stop Bits 1 4 Enable the CAPS LOCK key on your PC 5 This test verifies that the scale is transmitt...

Страница 13: ...necessary Try re calibrating the scale if this doesn t solve the problem Otherwise possible load cell damage due to overloading LO Warmup Mode The scale has been turned on without the platter install...

Отзывы: