APPENDIX C: Displayed Error Codes

CODE

MODE

MEANING / POSSIBLE SOLUTION

Normal Operating Mode

A  weight  greater  than  102%  of  the  scale’s  capacity  has

been  applied  to  the  scale.  Remove  the  weight  from  the

platter. Try re-calibrating the scale if this doesn’t  solve  the

problem.  Otherwise,  possible  load  cell  damage  due  to

overloading.

_ _ _ _ _ _

Normal Operating Mode

Underrange condition. Platter has been removed. Re-install

platter. Try re-calibrating the scale if this doesn’t  solve  the

problem.  Otherwise,  possible  load  cell  damage  due  to

shock loading.

– HI –

Warmup Mode

The scale has been turned on with a weight already present

on the platter. Remove the weight and then press the 

ZERO

key if necessary. Try re-calibrating the scale if this doesn’t

solve  the  problem.  Otherwise,  possible  load  cell  damage

due to overloading.

– LO –

Warmup Mode

The scale has been turned on without the platter installed.

Install  the  platter  and  then  press  the 

ZERO

  key  if  neces-

sary.  Try  re-calibrating  the  scale  if  this  doesn’t  solve  the

problem.  Otherwise,  possible  load  cell  damage  due  to

shock loading.

Err 0

Calibration Mode

Keyed-in weight value in Calibration Mode is less than 1.5

lb (0.7 kg). Use a larger test weight or re-adjust value.

Err 1

Calibration Mode

Keyed-in  weight  value  in  Calibration  Mode  is  larger  than

150 lb (70 kg). Use a smaller test weight or re-adjust value.

Err 2

Calibration Mode

Internal  resolution  is  not  high  enough  to  process  keyed-in

weight  value  in  Calibration  Mode.  Verify  test  weight  and

value.

Err 3

All Modes

Diagnostics check error - EEPROM Read

Err 4

All Modes

Diagnostics check error - EEPROM Write

Содержание PC 150

Страница 1: ...on 1 2 September 14 2000 Contents subject to change without notice Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont MN 56031 Tel 800 637 0529 Tel 507 238 8702 Fax 507 238 8271 E mail s...

Страница 2: ...l communi cation port The scale can be configured to display its divisions in two separate modes non NTEP mode and NTEP mode The scale is shipped from the factory configured in non NTEP mode The NTEP...

Страница 3: ...The scale is now ready for use ON OFF SWITCH DC POWER JACK RS 232 PORT ID TAG CALIBRATION SWITCH COVER PC 150 Back Panel The female 9 Pin RS 232 connector is used to interface to a PC Shown at right a...

Страница 4: ...the displayed weight is within 0 25 divisions of true zero NET Indicates that the scale is displaying net weight GROSS Indicates that the scale is displaying gross weight TARE Indicates that the tare...

Страница 5: ...S Toggles between Gross and Net weight display if a tare has been established TARE Used to establish the weight on the platform as the tare weight You cannot establish zero or negative gross weight as...

Страница 6: ...off to the unit 2 While holding down the lb kg key turn the power back on 3 When the display shows A1 the unit is in User Menu mode and you can release the lb kg key Shown at right are the direc tion...

Страница 7: ...Serial Data Format Selects the data format to be transmitted via the serial port to a printer or computer 0 Consolidated Controls Format 1 Toledo 8213 Format 2 NCI 3825Format 3 Transcell Technology F...

Страница 8: ...Shown at right are the di rectional and SET key assign ments SET SETUP MENU CHART F1 Display Mode F6 Zero Range F9 Calib Unit 0 1 F11 Display Check F12 Calibration Press ZERO key to begin Press ZERO...

Страница 9: ...one level begins the test sequence Press ZERO key to begin sequence F12 Calibration Places scale into the calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZER...

Страница 10: ...y pressing the PRINT key or the TARE key 5 After setting the exact value press the NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then freez...

Страница 11: ...0 6 7 Segment Green LED OVER CAPACITY ANNUNCIATION 102 of Full Scale Capacity OPERATING TEMPERATURE RANGE 32 F to 104 F 0 C to 40 C PLATTER SIZE 13 x 13 330 mm x 330 mm POWER SOURCE AC Adapter 12VDC 5...

Страница 12: ...al Select the proper settings for the following using the scale s settings Baud Rate Data Bits Parity Stop Bits 1 4 Enable the CAPS LOCK key on your PC 5 This test verifies that the scale is transmitt...

Страница 13: ...necessary Try re calibrating the scale if this doesn t solve the problem Otherwise possible load cell damage due to overloading LO Warmup Mode The scale has been turned on without the platter install...

Отзывы: