Page 8 - 1
CHAPTER 8: LEGAL FOR TRADE SEALING
8.1
STAINLESS STEEL ENCLOSURE
The 400ES indicator in the stainless steel enclosure can be sealed for commercial (Legal
for Trade) applications as follows.
1.
Power off the indicator.
2.
On the rear cover of the indicator, locate the calibration switch cover.
3.
Thread a wire security seal through both drilled head screws securing the calibration
switch cover as well as the two drilled head screws holding on the rear panel.
Older units
1.
Power off the indicator.
2.
Locate the two adjacent drilled head screws securing the rear cover.
3.
Thread a wire security seal through two drilled head screws securing the rear cover.