S A G E M E T E R I N G , I N C .
O p e r a t i o n s a n d I n s t r u c t i o n M a n u a l
64
REV. 10-SIX/SRX
HART Wiring Instructions
DC INPUT POWER WIRING
1. Turn off power source.
2. Open terminal block cover.
3. Input Power connections:
a. Requires maximim of 2.4 watts
b. Connect the positive lead (+) to terminal B5
and the Negative lead (–) to B6
–
+
INPUT POWER SUPPLY
VOLTAGE 24 VDC – 23.4 WATTS
B5 – POSITIVE (+)
B6 – NEGATIVE (–)
Loop
Power
Supply
HART field
calibrator
4. Connect 4-20 mA:
a. Connect isolated positive (+) lead to B4
b. Connect isolated negative (–) lead to C5
5. Connect a HART field calibrator as shown
insuring that there is a minimum of 250 ohm
resistance in the top loop.
6. For Pulsed output connect B4 (+) to C4 (–).
B5
B6
250–500
Ω
B4
C5