25
Upon coming across any problem in operation, follow the instructions below to check and
repair the problem, and/or get in touch with your representative.
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS
Imprevistos
Provável Causa
Soluções
Imprevist
Probable Cause
Solution
- The Curing Light does not
work.
- Battery in the handpiece
without charge.
- Overheating protection
activated “error code:
oU
.
- LED damaged “error
code:
Er
.
- Recharge the handpiece on
the base for 3 hours.
- Wait a few minutes.
- Get in touch with the Saevo
technical assistance.
- The equipment is not
polymerizing the resins.
- Resin not appropriate for the
wavelength range of the LED
curing lights.
- Light conductor fastened
incorrectly.
- Residue of resin in the light
conductor.
- Light conductor with protec-
tion cover.
- Acquire the appropriate re-
sin for the wavelength of the
curing light, i.e., which contain
photoinitiators with canforo-
quinone.
- Fasten the light conductor
correctly.
- Clean the light conductor.
- Remove the protection cover
from the light conductor.
- Inadequate luminous power. - Light conductor fastened
incorrectly.
- Problems with the light con-
ductor.
- Reduced capacity of the
battery.
- Fasten the light conductor
correctly.
- Replace the light conductor.
- Get in touch with the Saevo
technical assistance.