
USER MANUAL - SIP POWER ION PUMP CONTROLLER
M.HIST.0109.23
Rev.1
Page
32
of
62
9.3.
Registers Detailed Description
A detailed description of the available registers follows.
Manufacturing Data Registers
These registers are useful to retrieve manufacturing information.
CARD_TYPE – 0x1000 – Card Type
Bit 15 14 13 12 11 10 9 8
‒
‒
‒
‒
‒
‒
‒
‒
R/W
R R R R R R R R
Bit 7 6 5 4 3 2 1 0
‒
‒
‒
‒
‒
‒
ETHERNET
DISPLAY
R/W
R R R R R R R R
Bit 15:2 – Reserved bits
These bits are reserved for future use.
Bit 1 – Ethernet flag
If this bit is set, it means that Sip Power features a Ethernet connection and network registers are
addressable.
Bit 1 – Display flag
If this bit is set, it means that Sip Power features a display on the front panel.
HW_CODE – 0x1001 – Hardware Code
Bit 15 14 13 12 11 10 9 8
HARDWARE REVISION - MAJOR NUMBER
R/W
R R R R R R R R
Bit 7 6 5 4 3 2 1 0
HARDWARE REVISION - MINOR NUMBER
R/W
R R R R R R R R
Bit 15:8 – Hardware revision major number
This is the hardware revision major number.
Bit 7:0 – Hardware revision minor number
This is the hardware revision minor number.
SW_VERSION – 0x1002 – Software Version
Bit 15 14 13 12 11 10 9 8
SOFTWARE VERSION - MAJOR NUMBER
R/W
R R R R R R R R
Bit 7 6 5 4 3 2 1 0
SOFTWARE VERSION - MINOR NUMBER
R/W
R R R R R R R R
Содержание SIP POWER
Страница 54: ...USER MANUAL SIP POWER ION PUMP CONTROLLER M HIST 0109 23 Rev 1 Page 54 of 62 13 DECLARATION OF CE CONFORMITY...
Страница 55: ...USER MANUAL SIP POWER ION PUMP CONTROLLER M HIST 0109 23 Rev 1 Page 55 of 62...
Страница 62: ...USER MANUAL SIP POWER ION PUMP CONTROLLER M HIST 0109 23 Rev 1 Page 62 of 62...