USER MANUAL - SIP POWER ION PUMP CONTROLLER
M.HIST.0109.23
Rev.1
Page
19
of
62
Second character: setting and status of
SW1
switch
In Stop status the displayed character indicates the selected setting.
A white space means
SW1
is OFF, while “S” means
SW1
is ON.
In Active status the symbol indicates the current status of the switch (as set on the connector
IN/OUT INTEFACE of back panel, see wiring diagram for details):
= open switch and = closed switch
Third character: setting and status of
SW2
switch
In Stop status the displayed character indicates the selected setting.
A white space means
SW2
is OFF, while “S” means that
SW2
is ON and set to Simple working
mode and “W” means that
SW2
is on and set to Window working mode.
In Active status the symbol indicates the current status of the switch (as set on the connector
IN/OUT INTEFACE of back panel, see wiring diagram for details):
= open switch and = closed switch
Fourth character: setting and status of
SW3
switch
In Stop status the displayed character indicates the selected setting.
A white space means
SW3
is OFF, while “S” means that
SW3
is ON and set to Simple working
mode and “W” means that
SW3
is on and set to Window working mode.
In Active status the symbol indicates the current status of the switch (as set on the connector
IN/OUT INTEFACE of back panel, see wiring diagram for details):
= open switch and = closed switch
Supply time
This field refers to the IP working time since the high voltage supply starts. For short times (less
than 1 hour) the value is shown in minutes (“m”), for longer times in hours (“h”).
Current value of electrical variables
On the second row the operating values are shown.
The current flowing through the ion pump is shown on the left side of the display. The pump
voltage, output power and pump current can be alternatively displayed by pushing the pushbutton
A / V / W
.
The pump voltage is shown in V (Volt), current measurement automatically sets to nA - uA - mA
(nano (10
-9
) - micro (10
-6
) - milli (10
-3
) - Ampère) with a resolution of about 10nA, power range is in
mW or W (milli Watt (10
-3
W) or Watt)
Note: the voltage here reported is the actual value measured on the output in real time, without any
average, and without any processing. So it is possible that it changes slightly instantly.