Chapter 1: Introduction 

 

5

 

Chapter 1: Introduction 

 

 

General information  

The  BP4  blood  pressure  monitor  uses  the  oscillometric  method  for measuring  Non-Invasive  Blood  Pressure  (NIBP). 
NIBP measurement is based on the principle that pulsatile blood flow through an artery creates oscillations of the arterial 
wall. The oscillometric device uses a blood pressure cuff to sense these oscillations that appear as tiny pulsations in cuff 
pressure. The oscillometric devices measure the amplitude of pressure changes in the occluding cuff as the cuff deflates 
from above systolic pressure. The amplitude suddenly increases as the pulse breaks through the occlusion in the artery. 
As the cuff pressure decreases further, the pulsations increase in amplitude, reach a maximum (which approximates to 
the max amplitude pressure (MAP)), and then diminish. The oscillometric method measures the MAP and determines 
the systolic and diastolic pressures. 
In this monitor, we have IHB technology (IHB = Irregular heartbeat). The IHB technology analyses pulse frequency and 
detects irregular heartbeat during blood pressure measurement. If an irregular heartbeat occurs during the measurement, 
the IHB symbol is displayed. IHB does not replace a cardiac examination, but it serves to detect pulse irregularities at an 
early  stage.  This  is  automatically  done  by  the  device  without  the  need  for  any  adjustments.  Unlike  most  common 
monitors,  BP4  monitor  can  distinguish  between  arrhythmias  and  artefacts  due  to  great  technical  and  physiological 
precision in measurements. 
The device estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure (approximately 
30-40 mmHg above the patient’s systolic pressure).

 

The advantage is Personalized Inflation for maximum comfort. 

This  user  manual  contains  important  safety  information  and  provides  step-by-step  instructions  for  using  the  blood 
pressure monitor. Read this information carefully before you use the device and save it for future reference.

 

 

Benefits 

 

Ability to measure in different pressure ranges.  

 

Ability to measure in dialysis patients, pregnant women.  

 

Ability to measure in low gains.  

 

Automatic saving of blood pressure measurements (60 records per user). 

 

Simple User-Interface - 4 button operation & large display. 

 

Inflation level is automatically set. 

 

Working with batteries (4 x 1.5V “AA” batteries or rechargeable lithium batteries) or AC adapter.  

 

IHB technology - Irregular Heartbeat Technology analyses pulse frequency and detects irregular heartbeat during 
blood pressure measurement. 

 

Comfortable cuff for all common arm sizes. 

 

Cuff holder & handgrip for carrying.  

 

Medical Purpose 

The  device  is  a  digital  monitor  intended  for  use  in  measuring  blood  pressure  and  pulse  rate  in  adult  and 
pediatric patient population with arm circumference ranging from 21 cm to 55 cm (from 8 inches to 22 inches). 

 

Intended User 

This device should be used by a medical professional. 

 

Patient Population 

This device is intended for use on adults and children of age 3-years and older. 

 

Environment 

The instrument is designed for use in physicians' offices, hospitals, clinics and other medical facilities.  
 

Measurement Parameter 

 

Non-Invasive Blood Pressure 

 

Pulse rate 

 

 

Содержание BP4

Страница 1: ...User Manual Digital Blood Pressure Monitor Model BP4 D01027 V1 Software version 220 1 x x POOYANDEGAN RAH SAADAT Co...

Страница 2: ...21 77960719 98 21 77962181 Fax 98 21 77964239 Customer Services Tel 98 21 73098000 98 21 77798910 Cell 98 912 1977157 Fax 98 21 77960761 Legal responsible Trionara Technologies AB Polygonv gen 21 1876...

Страница 3: ...8 Installation of Batteries for devices which have no built in rechargeable batteries 8 Installation of rechargeable lithium Batteries 8 Using the AC Adapter 8 Symbols of Display 10 Chapter 2 System...

Страница 4: ...the used expressions in this manual Version Information This manual has a version number The version number changes whenever the manual is updated due to technical specification changes The version in...

Страница 5: ...r manual of the device and pay attention to the warnings and cautions This symbol indicates that the equipment shall be disposed of in an environmentally friendly manner Direct current CE mark Serial...

Страница 6: ...xtent Do not use an AC adapter or battery pack not specified for this unit Do not place heavy objects on the AC adapter cable or allow the unit to sit on the cord Do not use a broken power cord or AC...

Страница 7: ...ous cuff pressure due to connection tubing kinking may cause blood flow interference and result in harmful injury to the patient NIBP reading can be affected by the measurement site the position of th...

Страница 8: ...nd physiological precision in measurements The device estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure approximately 30 40 mmHg above the patient s systoli...

Страница 9: ...ted and the results showed a reliable measurement of the NIBP module in theses patient The performance of NIBP module in pregnant women was evaluated and the results showed a reliable measurement of t...

Страница 10: ...tor Front of the monitor a Display b Next Increase c Mode d Previous Decrease e Charging LED f START STOP button Left side right side and bottom of the monitor g AC adapter jack left side h Battery co...

Страница 11: ...connector and replace with a new one 3 Install the battery cover and fasten it with both screws Be careful not to pinch the lead wire 4 Connect the main unit to the AC adapter to charge the new batte...

Страница 12: ...n where it is difficult to plug and unplug the AC adapter We recommend keeping batteries in your monitor at all times even if you choose to use the AC adapter If only the AC adapter is used without ke...

Страница 13: ...itchable Systolic pressure Diastolic pressure Error number Heartbeat Measurement unit Measurement unit Bluetooth indicator Memory indicator Record number Appear when viewing stored readings Time Date...

Страница 14: ...he month number setting mode The month number is displayed as a blink and changes by pressing the up or down keys By pressing the key again the day number setting mode is activated The day number is d...

Страница 15: ...ff is used the measured blood pressure value tends to be lower than the actual blood pressure If too small a cuff is used the measured blood pressure value tends to be higher Before and during measure...

Страница 16: ...e correct arm position and is at the same height as the heart Start measurement 1 Press the START STOP button once to switch on the device 2 Attach the cuff to your arm see Attaching the cuff and make...

Страница 17: ...e used for disinfection Note Take extra care when cleaning the screen of the monitor because it is more sensitive to rough cleaning methods than the housing Don t spray a liquid directly on the screen...

Страница 18: ...Measuring pressure is more than upper limit 255mmHg for adult Repeat measurement If the error is displayed again use another method to measure the patient s blood pressure 11 EXCESSIVE MOTION Arm move...

Страница 19: ...uracy 2 or 2 BPM whichever is greater Initial cuff pressure 160 mmHg Safety Regulatory Standards IEC 60601 1 EN 60601 1 2 EN 60601 1 11 IEC 80601 2 30 ISO 81060 2 ISO 10993 5 ISO 10993 10 NIBP Accurac...

Страница 20: ...Appendix 1 Accessories 17 Appendix 1 Accessories PART Number NIBP Accessory P48028 NIBP Reusable Cuff Adult Single M5134 Tube Length 50 cm Limb Circumference 25 35cm...

Страница 21: ...t In case adjacent or stacked use is necessary the medical device should be observed to verify normal operation in the configuration in which it will be used During use or measurement portable RF comm...

Страница 22: ...should be at least 30 8 kV contact 2 kV 4 kV 8 kV 15 kV air N A Enclosure Electrostatic discharge ESD IEC 61000 4 2 Patient coupling Signal I O parts N A Input a c power Electrical fast transient bur...

Страница 23: ...patient monitor should assure that it is used in such an environment Electromagnetic Environment Guidance Compliance Level Port Immunity Test N A Input a c power Conducted RF IEC 61000 4 6 Patient cou...

Страница 24: ...Hz GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 800 960 810 870 930 28 0 3 2 Pulse modulation b 217 Hz GSM 1800 CDMA 1900 DECT LTE Band 1 3 4 25 UMTS 1700 1990 1720 1845 1970 28 0 3 2 Pulse modu...

Страница 25: ......

Отзывы: