Chapter 3: Blood Pressure Measurement   

12

 

Chapter 3: Blood Pressure Measurement 

 

 

Tips for proper measurement 

 
 

 

Note 

 

Before  using  the  unit,  make  sure  that  no  accessories  are  missing  and  that  the  unit  and 
accessories are not damaged. If an accessory is missing or there is damage, please contact your 
local representative

 

Attaching the cuff 

1.

 

Selecting the cuff 

o

 

Measure  the  circumference  of  the  patient's  arm  and  select  the  cuff  size  that  is  appropriate  for  the 
circumference. Select the cuff that is suitable for the patient from the cuffs exist in appendix1.  

2.

 

Connecting the cuff 

o

 

Connect the cuff tube to the NIBP connector on the unit and turn clockwise to lock 

 

Note 

 

The device can be used on either the right or left arm. Wrap the cuff on a bare arm or over thin 
clothing.  Thick  clothing  or  a  rolled  up  sleeve  will  cause  inaccurate  blood  pressure 
measurements.

 

 

Warning  

 

NIBP measurement should be performed on the upper arm. 

 

During NIBP, stop excessive body movement by the patient and minimize trembling. 

 

Use the appropriate cuff size to ensure correct measurements. If too large a cuff is used, the 
measured blood pressure value tends to be lower than the actual blood pressure. If too small 
a cuff is used, the measured blood pressure value tends to be higher. 

 

Before and during measurement, verify that none of the following apply to the patient: 

o

 

An inappropriate cuff size is used. 

o

 

The  part  where  the  cuff  is  wrapped  is  at  a  different  height  than  the  heart.  (A 
difference of 10 cm (4 inches) in height may cause a variation in the blood pressure 
value of up to 7 or 8 mmHg.) 

o

 

Body movement or conversing during measurement. 

o

 

Cuff wrapped over thick clothing. 

o

 

Pressure on the arm due to a rolled up sleeve. 

 

In the case of a cuff for adults, the cuff should be wrapped to a tightness that allows two 
fingers to be inserted in between the cuff and the arm. 

 

Do not use the cuff if it is damaged or has holes. 

 

Only company-approved

 

CUFF can be used with this device. The use of any other cuff may 

result in incorrect measurement. 

 

To  measure  correctly,  it  is  recommended  that  the  patient  relax  and  not  talk  during 
measurement. 

To measure correctly, it is recommended that the patient rest quietly for 5 minutes before 
measurement. In the case of a cuff for adults, the cuff should be wrapped to a tightness that 
allows two fingers to be inserted in between the cuff and the arm. 

Содержание BP4

Страница 1: ...User Manual Digital Blood Pressure Monitor Model BP4 D01027 V1 Software version 220 1 x x POOYANDEGAN RAH SAADAT Co...

Страница 2: ...21 77960719 98 21 77962181 Fax 98 21 77964239 Customer Services Tel 98 21 73098000 98 21 77798910 Cell 98 912 1977157 Fax 98 21 77960761 Legal responsible Trionara Technologies AB Polygonv gen 21 1876...

Страница 3: ...8 Installation of Batteries for devices which have no built in rechargeable batteries 8 Installation of rechargeable lithium Batteries 8 Using the AC Adapter 8 Symbols of Display 10 Chapter 2 System...

Страница 4: ...the used expressions in this manual Version Information This manual has a version number The version number changes whenever the manual is updated due to technical specification changes The version in...

Страница 5: ...r manual of the device and pay attention to the warnings and cautions This symbol indicates that the equipment shall be disposed of in an environmentally friendly manner Direct current CE mark Serial...

Страница 6: ...xtent Do not use an AC adapter or battery pack not specified for this unit Do not place heavy objects on the AC adapter cable or allow the unit to sit on the cord Do not use a broken power cord or AC...

Страница 7: ...ous cuff pressure due to connection tubing kinking may cause blood flow interference and result in harmful injury to the patient NIBP reading can be affected by the measurement site the position of th...

Страница 8: ...nd physiological precision in measurements The device estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure approximately 30 40 mmHg above the patient s systoli...

Страница 9: ...ted and the results showed a reliable measurement of the NIBP module in theses patient The performance of NIBP module in pregnant women was evaluated and the results showed a reliable measurement of t...

Страница 10: ...tor Front of the monitor a Display b Next Increase c Mode d Previous Decrease e Charging LED f START STOP button Left side right side and bottom of the monitor g AC adapter jack left side h Battery co...

Страница 11: ...connector and replace with a new one 3 Install the battery cover and fasten it with both screws Be careful not to pinch the lead wire 4 Connect the main unit to the AC adapter to charge the new batte...

Страница 12: ...n where it is difficult to plug and unplug the AC adapter We recommend keeping batteries in your monitor at all times even if you choose to use the AC adapter If only the AC adapter is used without ke...

Страница 13: ...itchable Systolic pressure Diastolic pressure Error number Heartbeat Measurement unit Measurement unit Bluetooth indicator Memory indicator Record number Appear when viewing stored readings Time Date...

Страница 14: ...he month number setting mode The month number is displayed as a blink and changes by pressing the up or down keys By pressing the key again the day number setting mode is activated The day number is d...

Страница 15: ...ff is used the measured blood pressure value tends to be lower than the actual blood pressure If too small a cuff is used the measured blood pressure value tends to be higher Before and during measure...

Страница 16: ...e correct arm position and is at the same height as the heart Start measurement 1 Press the START STOP button once to switch on the device 2 Attach the cuff to your arm see Attaching the cuff and make...

Страница 17: ...e used for disinfection Note Take extra care when cleaning the screen of the monitor because it is more sensitive to rough cleaning methods than the housing Don t spray a liquid directly on the screen...

Страница 18: ...Measuring pressure is more than upper limit 255mmHg for adult Repeat measurement If the error is displayed again use another method to measure the patient s blood pressure 11 EXCESSIVE MOTION Arm move...

Страница 19: ...uracy 2 or 2 BPM whichever is greater Initial cuff pressure 160 mmHg Safety Regulatory Standards IEC 60601 1 EN 60601 1 2 EN 60601 1 11 IEC 80601 2 30 ISO 81060 2 ISO 10993 5 ISO 10993 10 NIBP Accurac...

Страница 20: ...Appendix 1 Accessories 17 Appendix 1 Accessories PART Number NIBP Accessory P48028 NIBP Reusable Cuff Adult Single M5134 Tube Length 50 cm Limb Circumference 25 35cm...

Страница 21: ...t In case adjacent or stacked use is necessary the medical device should be observed to verify normal operation in the configuration in which it will be used During use or measurement portable RF comm...

Страница 22: ...should be at least 30 8 kV contact 2 kV 4 kV 8 kV 15 kV air N A Enclosure Electrostatic discharge ESD IEC 61000 4 2 Patient coupling Signal I O parts N A Input a c power Electrical fast transient bur...

Страница 23: ...patient monitor should assure that it is used in such an environment Electromagnetic Environment Guidance Compliance Level Port Immunity Test N A Input a c power Conducted RF IEC 61000 4 6 Patient cou...

Страница 24: ...Hz GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 800 960 810 870 930 28 0 3 2 Pulse modulation b 217 Hz GSM 1800 CDMA 1900 DECT LTE Band 1 3 4 25 UMTS 1700 1990 1720 1845 1970 28 0 3 2 Pulse modu...

Страница 25: ......

Отзывы: