Chapter 4: Care and Cleaning  

14

 

Chapter 4: Care and Cleaning

 

 

 

Cleaning and disinfecting should be performed in accordance with your

 

facility's infection control practice.

 

Wipe with a 

cloth that has been moistened with isopropyl alcohol diluted to

 

50 v/v%, or ethyl alcohol (disinfection alcohol) diluted 

to 76.9 - 81.4 v/v%

 

and wrung out.

 

Do not wipe the Power connector or allow it to become wet.

 

Use a moistened cotton 

bud to remove dust that has accumulated on the

 

vent ports.

 

The device requires no routine service other than cleaning, 

and visually

 

checking the cuffs, tubing, etc. 

 

Note 

 

Do  not  sterilize  by  autoclave  or  gas  sterilization  (EOG,  formaldehyde  gas,  high-
concentration ozone, etc.). 

 

If using an antiseptic solution for cleaning, follow the instructions of the manufacturer. 

 

 

 

Display

 

screen care

 

In-between patients and as required use clean and soft cloth with screen cleaner or mild soapy water and with 
Isopropyl alcohol may be used for disinfection. 
 

Note 

 

Take extra care when cleaning the screen of the monitor because it is more sensitive to rough 
cleaning methods than the housing.  

 

Don't spray a liquid directly on the screen.  

 

 

Accessory care

 

Cuff/cuff tube:  
Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v%

 

dilution of isopropyl alcohol, or a 

76.9 to 81.4 v/v% dilution of disinfection

 

ethanol (ethyl alcohol). Do not allow any liquids inside the cuff. If a 

liquid gets in the cuff, dry the i

t.  

 

Warning  

 

This device is not washable. Never immerse the device in water and do not rinse it under 
running water. 

 

Never  use  compressed  air,  scouring  pads,  abrasive  cleaning  agents  or  aggressive  liquids 
such as petrol or acetone to clean the device. 

 

Sterilization  may  cause  damage  to  the  device  and  is  therefore  not  recommended  for  this 
device otherwise indicated in the instructions delivered with accessories or your hospital’s 
servicing schedule. 

 

Do not clean this unit with gasoline, paint thinner or high concentration alcohol. 

 

Do  not  disinfect  this  unit  by  autoclave  or  gas  sterilization  (ETO,  glutaraldehyde  or  high 
concentration ozone).  

 
 

Содержание BP4

Страница 1: ...User Manual Digital Blood Pressure Monitor Model BP4 D01027 V1 Software version 220 1 x x POOYANDEGAN RAH SAADAT Co...

Страница 2: ...21 77960719 98 21 77962181 Fax 98 21 77964239 Customer Services Tel 98 21 73098000 98 21 77798910 Cell 98 912 1977157 Fax 98 21 77960761 Legal responsible Trionara Technologies AB Polygonv gen 21 1876...

Страница 3: ...8 Installation of Batteries for devices which have no built in rechargeable batteries 8 Installation of rechargeable lithium Batteries 8 Using the AC Adapter 8 Symbols of Display 10 Chapter 2 System...

Страница 4: ...the used expressions in this manual Version Information This manual has a version number The version number changes whenever the manual is updated due to technical specification changes The version in...

Страница 5: ...r manual of the device and pay attention to the warnings and cautions This symbol indicates that the equipment shall be disposed of in an environmentally friendly manner Direct current CE mark Serial...

Страница 6: ...xtent Do not use an AC adapter or battery pack not specified for this unit Do not place heavy objects on the AC adapter cable or allow the unit to sit on the cord Do not use a broken power cord or AC...

Страница 7: ...ous cuff pressure due to connection tubing kinking may cause blood flow interference and result in harmful injury to the patient NIBP reading can be affected by the measurement site the position of th...

Страница 8: ...nd physiological precision in measurements The device estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure approximately 30 40 mmHg above the patient s systoli...

Страница 9: ...ted and the results showed a reliable measurement of the NIBP module in theses patient The performance of NIBP module in pregnant women was evaluated and the results showed a reliable measurement of t...

Страница 10: ...tor Front of the monitor a Display b Next Increase c Mode d Previous Decrease e Charging LED f START STOP button Left side right side and bottom of the monitor g AC adapter jack left side h Battery co...

Страница 11: ...connector and replace with a new one 3 Install the battery cover and fasten it with both screws Be careful not to pinch the lead wire 4 Connect the main unit to the AC adapter to charge the new batte...

Страница 12: ...n where it is difficult to plug and unplug the AC adapter We recommend keeping batteries in your monitor at all times even if you choose to use the AC adapter If only the AC adapter is used without ke...

Страница 13: ...itchable Systolic pressure Diastolic pressure Error number Heartbeat Measurement unit Measurement unit Bluetooth indicator Memory indicator Record number Appear when viewing stored readings Time Date...

Страница 14: ...he month number setting mode The month number is displayed as a blink and changes by pressing the up or down keys By pressing the key again the day number setting mode is activated The day number is d...

Страница 15: ...ff is used the measured blood pressure value tends to be lower than the actual blood pressure If too small a cuff is used the measured blood pressure value tends to be higher Before and during measure...

Страница 16: ...e correct arm position and is at the same height as the heart Start measurement 1 Press the START STOP button once to switch on the device 2 Attach the cuff to your arm see Attaching the cuff and make...

Страница 17: ...e used for disinfection Note Take extra care when cleaning the screen of the monitor because it is more sensitive to rough cleaning methods than the housing Don t spray a liquid directly on the screen...

Страница 18: ...Measuring pressure is more than upper limit 255mmHg for adult Repeat measurement If the error is displayed again use another method to measure the patient s blood pressure 11 EXCESSIVE MOTION Arm move...

Страница 19: ...uracy 2 or 2 BPM whichever is greater Initial cuff pressure 160 mmHg Safety Regulatory Standards IEC 60601 1 EN 60601 1 2 EN 60601 1 11 IEC 80601 2 30 ISO 81060 2 ISO 10993 5 ISO 10993 10 NIBP Accurac...

Страница 20: ...Appendix 1 Accessories 17 Appendix 1 Accessories PART Number NIBP Accessory P48028 NIBP Reusable Cuff Adult Single M5134 Tube Length 50 cm Limb Circumference 25 35cm...

Страница 21: ...t In case adjacent or stacked use is necessary the medical device should be observed to verify normal operation in the configuration in which it will be used During use or measurement portable RF comm...

Страница 22: ...should be at least 30 8 kV contact 2 kV 4 kV 8 kV 15 kV air N A Enclosure Electrostatic discharge ESD IEC 61000 4 2 Patient coupling Signal I O parts N A Input a c power Electrical fast transient bur...

Страница 23: ...patient monitor should assure that it is used in such an environment Electromagnetic Environment Guidance Compliance Level Port Immunity Test N A Input a c power Conducted RF IEC 61000 4 6 Patient cou...

Страница 24: ...Hz GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 800 960 810 870 930 28 0 3 2 Pulse modulation b 217 Hz GSM 1800 CDMA 1900 DECT LTE Band 1 3 4 25 UMTS 1700 1990 1720 1845 1970 28 0 3 2 Pulse modu...

Страница 25: ......

Отзывы: