6
Main Frame -
The rectangular piece that is fixed to the wheels - the main frame is made up of several
components:
Stabilizing Arms -
Two arms pull out to level the lift and must always be fully extended when the lift is in
use.
WARNING: Failure to fully extend both stabilizer arms during transfer could cause lift movement
and may result in injury. Both stabilizing arms must be fully extended whenever the lift is
positioned for use. Do not retract stabilizing arms during transfer as the lift could move and may
result in injury.
Housing -
ABS plastic cover shields main frame components from weather. The housing needs to be in
place prior to installing mast assembly.
Handle -
Connects to Main Frame through openings in housing.
Lift Operator
- The LiftOperator unit controls all lift operations
identical to the user hand set. Three cables connect to the
bottom of the Control Box to enable operation of the lift. The
largest connector is for the hand control. Connector # 1 is for
the motor cable (red stripe). Connector # 2 is for the actuator
cable (green stripe).
Touchpad Control
–
The touchpad control arrow keys can be used in the
event that the hand control is out of reach or fails.
Simply press and hold the arrow for the desired action
for one movement at a time
– Up, Down, Left, Right
Releasing the Arrow stops movement.
Battery Level LED Indicators
–
The Battery Level LED Indicators show battery charge levels. The LED’s will illuminate when
either the touchpad control or the hand control is activated and will stay lit for 10 seconds. At
greater than 50% the LED glows Green, at less than 50% the LED glows Amber, and less than
25% the LED glows red and indicates the battery requires charging. If the less than 25% LED
glows red, do not operate the lift. Remove the battery and fully charge before use.
Service Required LED
–
The Service Required LED will illuminate when the battery pack has been removed/replaced 120
times - approximately once every 4 months. This indicates that required maintenance for the lift
must be completed. Please see the maintenance section of the Lift
Owner’s Manual for the
requirements. The Service Required LED can be reset by inserting a standard FAT-formatted
USB mass storage device (not included) into the USB Port on the bottom of the controller
. When
the service light is illuminated the lift must not be used until required maintenance is
performed.
Emergency Stop Button
–
In the event of an emergency, or if you need to stop lift movement immediately, pressing the
Emergency Stop Button (red button on the control) will stop all lift movement. At the same time an
audible alert will sound for 10 sec. then pause for 5 sec. and then the sequence will repeat until
the button is reset by turning it in the direction of the arrows on the button (clockwise). The
Emergency Alert LED will also flash Red when the button is pressed and will remain flashing until
the button is reset.
CONNECTOR 1
(MOTOR CABLE)
CONNECTOR 2
(ACTUATOR CABLE)