55
Doc. No.
:
P30001PE0101-07
Note
•
Do not use alcohol when cleaning the clear panel on the instrument door.
Instead, use water for those parts.
•
70% ethanol is a flammable liquid. If you are using 70% ethanol, ensure that there
are no flames nearby.
13.2.
D-Ring Maintenance
To ensure complete connection and to prevent leakage, we recommend
performing maintenance of the D-Rings once every two weeks.
1.
Wearing gloves, wipe away old grease using a paper or cloth.
2.
Slightly apply some silicon grease to your gloved finger.
3.
Wipe the silicon grease onto the surfaces of the D-Rings (black rubber
rings on the nozzles).
Tip Ejector
D-ring
Nozzle